GENTLEPOWER LUX CONTRA ANGLE 25LPR
Report
- Report Number
- 3003637274-2015-00025
- Event Type
- Malfunction
- Date Received
- September 21, 2015
- Date of Event
- August 31, 2015
- Report Date
- September 8, 2015
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EKX
- PMA / PMN Number
- K073478
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- DENTIST
Narratives
THE ANALYSIS OF THE HANDPIECE SHOWED THAT IT WAS RUNNING OUT OF SPECIFICATION. THIS MEANS THAT THE CURRENT CONSUMPTION DURING THE TEST RUN WAS HIGHER THAN SPECIFIED. ROOT CAUSE IS THAT THE INNER PARTS ARE RUNNING STIFF DUE TO WORN BEARINGS. THE HEAT UP WAS REPRODUCIBLE DURING THE TEST RUN. IT ALSO SHOWED THAT THE PUSH BUTTON HAD CIRCULAR WARE MARKS ON THE INNER SIDE. THIS SHOWS THAT IT HAS BEEN PRESSED WHILE HANDPIECE WAS RUNNING. THIS IS NORMALLY THE CASE WHEN IT IS USED TO LIFT THE CHEEK OR TONGUE AWAY. TO AVOID SUCH ISSUES THE USER INSTRUCTION CONTAINS SEVERAL WARNINGS AND NOTES: WARNING: HAZARDS FOR THE CARE PROVIDER AND THE PATIENT. IN THE CASE OF DAMAGE, IRREGULAR RUNNING NOISE, EXCESSIVE VIBRATION, UNTYPICAL WARMING OR WHEN THE CUTTER OR GRINDER CANNOT BE HELD. DO NOT USE FURTHER AND NOTIFY SERVICE. CAUTION: BURNING HAZARD FROM HOT INSTRUMENT HEAD OR HOT INSTRUMENTS COVER. IF THE INSTRUMENT OVERHEATS, BURNS MAY ARISE IN THE ORAL AREA. NEVER CONTACT SOFT TISSUE WITH THE INSTRUMENT HEAD OR INSTRUMENT COVER. THE FOLLOWING GUIDELINES MUST BE OBSERVED TO ENSURE SAVE USE OF THE ELECTRICALLY DRIVEN CONTRA-ANGLE HANDPIECES: THE SERVICE INSTRUCTIONS FOR CONTRA-ANGLE HANDPIECES MUST BE PRECISELY FOLLOWING WHEN USING KAVO SPRAY OR QUATTROCARE CARE SYSTEMS. BEFORE EACH USE, THE CONTRA-ANGLE HANDPIECE MUST BE CHECKED FOR EXTERNAL DAMAGE. BEFORE EACH USE, PERFORM A TEST RUN WITH THE CONTRA-ANGLE HANDPIECE, AND WATCH FOR ATYPICAL HEATING AND UNUSUAL NOISE AND VIBRATION. IMMEDIATELY STOP USING CONTRA-ANGLE HANDPIECES THAT ACT UNUSUAL. NEVER PRESS THE PUSH BUTTON DURING OPERATION. THIS ALSO INCLUDES LIFTING THE CHEEK OR TONGUE! TO ENSURE PROPER FUNCTION, THE MEDICAL DEVICE MUST BE SET UP ACCORDING TO THE REPROCESSING METHODS DESCRIBED IN THE KAVO INSTRUCTIONS FOR USE, AND THE CARE PRODUCTS AND CARE SYSTEMS DESCRIBED THEREIN MUST BE USED. KAVO RECOMMENDS SPECIFYING A SERVICE INTERVAL AT THE DENTAL OFFICE FOR A LICENSED SHOP TO CLEAN, SERVICE AND CHECK THE FUNCTIONING OF THE MEDICAL DEVICE. THIS SERVICE INTERVAL DEPENDS ON THE FREQUENCY OF USE AND SHOULD BE ADJUSTED ACCORDINGLY. (B)(4).
DURING A STANDARD DENTAL TREATMENT THE HANDPIECE HEATED UP AND BURNED THE PATIENT ON INNER CHEEK. THERE WAS NO MEDICAL CARE NECESSARY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 622311 | GENTLEPOWER LUX CONTRA ANGLE 25LPR | ELECTRICAL DENTAL HANDPIECE | EKX | KAVO DENTAL GMBH | 25 LPR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |