FDA Adverse Event Malfunction Summary report: N

3612 LITE GLOVE 2X80

MDR report key: 5090681 · Received September 21, 2015

Report

Report Number
9612030-2015-00089
Event Type
Malfunction
Date Received
September 21, 2015
Report Date
September 4, 2015
Manufacturer
COVIDIEN
Product Code
FMC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN INVESTIGATION IS CURRENTLY UNDER WAY; UPON COMPLETION THE RESULTS WILL BE FORWARDED.

Additional Manufacturer Narrative · 1

THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ABNORMAL PROCESS CONDITION WAS PRESENT DURING THE MANUFACTURING OF THE PRODUCT THAT COULD HAVE LED TO THE REPORTED CONDITION. THE DHR REVIEW SHOWED THAT ALL ACCEPTANCE CRITERIA INSPECTIONS PER ESTABLISHED SAMPLING LEVELS WERE WITHIN ACCEPTABLE LIMITS DURING THE PRODUCTION PROCESS. THREE CASES WITH UNUSED PRODUCT AND LOT NUMBER 5138101664X WERE RECEIVED FOR EVALUATION. THE SAMPLES WERE VISUALLY INSPECTED ACCORDING TO THE PRODUCT SPECIFICATIONS AND NONE OF THE SAMPLES DEMONSTRATED THE REPORTED CONDITION. THE REPORTED CONDITION COULD NOT BE CONFIRMED HOWEVER, BASED ON PREVIOUS EVALUATIONS ON SAMPLES REPORTED WITH A SIMILAR CONDITION THE POTENTIAL ROOT CAUSE THAT WAS DETERMINED FOR THIS ISSUE WAS A PROBLEM THAT OCCURRED DURING THE FORMING PROCESS IN WHICH THE FORMING TOOL REACHES A HIGH TEMPERATURE. DUE TO PROPERTIES OF THE MATERIAL, THE HEAT IS CONCENTRATED AT THE POINT OF CONTACT WITH THE FILM AND NOT DISSIPATED THROUGHOUT THE REST OF THE TOOLING THEREFORE IT DOES NOT BALANCE OUT. THIS MAY CAUSE THE FILM TO OVERHEAT AND COULD POTENTIALLY AFFECT THE STRENGTH OF THE MATERIAL. AS A CORRECTIVE ACTION THE MATERIAL USED FOR THE PLUG ASSISTS WAS CHANGED; THE DIFFERENCE IN THIS CHANGE IS ITS PROPERTIES, THE HEAT STABILIZED AROUND THE TOOLING (NOT ONLY THE CONTACT POINT WITH THE FILM) HAVING A LOWER IMPACT ON THE FILM FORMING PROCESS. AN ENGINEERING TEST REQUEST (ETR) WAS PERFORMED TO DETERMINE THE BEST TEMPERATURES TO FORM THE LITE GLOVE WITH THE NEW FORMING TOOL USING THE NEWER, IMPROVED MATERIAL. AFTER THE ETR WAS APPROVED, VALIDATION WAS PERFORMING TO CONFIRM THE PROPOSED PARAMETERS AND ENSURE THE QUALITY OF THE PRODUCT. A FORMAL CORRECTIVE AND PREVENTATIVE ACTION WAS IMPLEMENTED AS A RESULT. IF ADDITIONAL INFORMATION IS RECEIVED WARRANTING FURTHER ANALYSIS, THE INVESTIGATION WILL BE RESUMED. THIS COMPLAINT WILL BE USED FOR TRACKING AND TRENDING PURPOSES.

Description of Event or Problem · 1

IT WAS REPORTED TO COVIDIEN ON (B)(6) 2015 THAT A CUSTOMER HAD AN ISSUE WITH A LITE GLOVE. THE CUSTOMER REPORTS THE MATERIAL OF SEVERAL LITE GLOVES RIPPED OPEN WHEN PUTTING THEM OVER THE LIGHT HANDLE. THE LITE GLOVES ARE TOO SMALL. THE ISSUE WAS NOTICED BEFORE THE PATIENT WAS IN THE OPERATING ROOM. THERE WAS NO PERSON INJURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
623734 3612 LITE GLOVE 2X80 LITE GLOVE FMC COVIDIEN 31140257 51438101664X

Patients

Seq Age Sex Outcome Treatment
1