FDA Adverse Event
Malfunction
Summary report: N
HEARTSINE SAMARITAN 300P AND PAD PAK
MDR report key: 5090472
·
Received September 21, 2015
Report
- Report Number
- 3004123209-2015-01227
- Event Type
- Malfunction
- Date Received
- September 21, 2015
- Date of Event
- September 14, 2015
- Report Date
- October 15, 2015
- Manufacturer
- HEARTSINE TECHNOLOGIES LTD
- Product Code
- MKJ
- PMA / PMN Number
- K041067
- Removal / Correction Number
- Z-0124-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PAD-PAK WAS FIRST INSTALLED SUCCESSFULLY ON THE (B)(6) 2012 AND PERFORMED TO SPECIFICATION UP TO THE (B)(6) 2014. MULTIPLE MANUAL POWER UPS OF MAINLY TEN MINUTE DURATION WERE RECORDED BETWEEN THE (B)(6) 2014 AND THE (B)(6) 2015. INVESTIGATION FOUND THE REPORTED FAULT CAN BE ATTRIBUTED TO MEMBRANE FAILURE. THE TEN MINUTE TIME OUTS WOULD SUGGEST A FAILING MEMBRANE. THE FAULT WAS WITNESSED DURING THE INVESTIGATION. THE FAULT COULD NOT BE REPLICATED WITH A NEW MEMBRANE FITTED. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
Description of Event or Problem · 1
THERE WAS NO PATIENT INVOLVED IN THIS EVENT. UNIT POWERS ON BY ITSELF WITHOUT MANUAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 623206 | HEARTSINE SAMARITAN 300P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES LTD | SAM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |