FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 5090472 · Received September 21, 2015

Report

Report Number
3004123209-2015-01227
Event Type
Malfunction
Date Received
September 21, 2015
Date of Event
September 14, 2015
Report Date
October 15, 2015
Manufacturer
HEARTSINE TECHNOLOGIES LTD
Product Code
MKJ
PMA / PMN Number
K041067
Removal / Correction Number
Z-0124-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PAD-PAK WAS FIRST INSTALLED SUCCESSFULLY ON THE (B)(6) 2012 AND PERFORMED TO SPECIFICATION UP TO THE (B)(6) 2014. MULTIPLE MANUAL POWER UPS OF MAINLY TEN MINUTE DURATION WERE RECORDED BETWEEN THE (B)(6) 2014 AND THE (B)(6) 2015. INVESTIGATION FOUND THE REPORTED FAULT CAN BE ATTRIBUTED TO MEMBRANE FAILURE. THE TEN MINUTE TIME OUTS WOULD SUGGEST A FAILING MEMBRANE. THE FAULT WAS WITNESSED DURING THE INVESTIGATION. THE FAULT COULD NOT BE REPLICATED WITH A NEW MEMBRANE FITTED. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.

Description of Event or Problem · 1

THERE WAS NO PATIENT INVOLVED IN THIS EVENT. UNIT POWERS ON BY ITSELF WITHOUT MANUAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
623206 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES LTD SAM