KARL STORZ
Report
- Report Number
- MW1030918
- Event Type
- Injury
- Date Received
- January 21, 2004
- Date of Event
- January 12, 2004
- Report Date
- January 21, 2004
- Manufacturer
- KARL STORZ ENDOSCOPY
- Product Code
- KQT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- ID, US
- Reporter Occupation
- RISK MANAGER
Narratives
DURING TRANSURETHRAL RESECTION OF THE PROSTATE PROCEDURE THE UROLOGY CONTINUOUS FLOW PUMP WAS CONNECTED INCORRECTLY RESULTING IN AIR BEING PUMPED INTO BLADDER RATHER THAN FLUID EVACUATED. THIS RESULTED IN BLADDER PERFORATION, PNEUMOPERITONEUM AND PNEUMOTHORAX, AND HEMODYNAMIC CHANGES LEADING TO CARDIAC ARREST WITH SUCCESSFUL RESUSCITATIVE EFFORTS. OPEN LAPAROTOMY WAS PERFORMED TO REPAIR THE BLADDER AND PT WAS TRANSFERRED TO INTENSIVE CARE FOR POST-OP MANAGEMENT. STAFF RE-EDUCATED REGARDING USE OF EQUIPMENT HOWEVER A CONCERN HAS BEEN IDENTIFIED WITH THE EQUIPMENT THAT ALLOWS INCORRECT CONNECTIONS AND REPORTER IS CONCERNED THAT THIS MAY BE AN EQUIPMENT DESIGN FLAW.
ADD'L INFO REC'D FROM MFR 3/19/2004: AS STATED IN THE REPORT, THE HOSPITAL'S BIO-MED HAS EVALUATED THE UNIT AND CONFIRMED IT WAS IN OPERATIONAL CONDITION AND THAT THEY DO NOT CONSIDER THERE WAS A NEED TO RETURN THE UNIT FOR EVALUATION. THIS HOSPITAL HAS HAD THE PUMP FOR OVER 12 YEARS, IT HAS NEVER BEEN RETURNED FOR REPAIR OR MAINTENANCE. WHEN THIS PRODUCT WAS FIRST INTRODUCED TO THE MARKET IN 1983 (K832009), IT HAD CLEAR DIRECTIONS OF "TO WASTE" AND "FROM INSTRUMENT" ON THE FACEPLATE. MFR THEN ADDED A PLASTIC PLATE OVER THE PUMP HEAD (DECEMBER 1991) WITH BRIGHT YELLOW CAUTION STATEMENT AND BIG ARROWS SHOWING THE DIRECTION OF THE FLOW. IN 2002, MFR ADDED ANOTHER SET OF BRIGHT YELLOW STICKERS TO THE FACEPLATE DUPLICATING THE DIRECTION OF THE FLOW, JUST IN CASE CUSTOMER DID NOT HAVE THE YELLOW PLATE ON.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | KARL STORZ | UROLOGY CONTINUOUS FLOW PUMP | KQT | KARL STORZ ENDOSCOPY | 27224P | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Hospitalization| L| R |