SPACELABS ARKON ANESTHESIA WORKSTATION

MDR report key: 5088911 · Received September 18, 2015

Report

Report Number
9611295-2015-00011
Event Type
Malfunction
Date Received
September 18, 2015
Date of Event
March 25, 2015
Report Date
May 27, 2020
Manufacturer
SPACELABS HEALTHCARE LTD.
Product Code
CBK
PMA / PMN Number
K113051
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 1

ON 18-SEP-2015: THE CUSTOMER REPORTED THE DEVICE PASSED MORNING CHECKOUT PROCEDURES PRIOR TO USE AND EMERGENCY BACKUP SYSTEMS WORKED AS DESIGNED DURING THE EVENT SO THERE WAS NO INTERRUPTION OF PATIENT CARE. THE MACHINE RETURNED TO NORMAL. SERVICE AFTER A SYSTEM REBOOT. SPACELABS HAS LAUNCHED AN INVESTIGATION OF THIS EVENT AND WILL FILE A SUPPLEMENTAL REPORT WHEN THE INVESTIGATION IS COMPLETE. ON 12-FEB-2016: BASED ON THE SERVICE LOGS AND ONSITE INVESTIGATION BY A SPACELABS FIELD SERVICE ENGINEER, FAULTY COMMUNICATION FROM THE DU TO ACB WAS FOUND. THE ISSUE WAS NOT REPRODUCIBLE; NO TROUBLE WAS FOUND UPON INVESTIGATION; THE UNIT PASSED ALL FUNCTIONAL TESTS. THE UNIT WAS RETURNED TO SERVICE WITHOUT FURTHER ISSUE. DURING THE EVENT CLINICAL STAFF RESPONDED APPROPRIATELY PER THEIR TRAINING; MANUAL VENTILATION, MEDICAL GASES AND ALARMS REMAINED OPERATIONAL. THIS INVESTIGATION IS CONSIDERED COMPLETE AND THIS PARTICULAR ISSUE CLOSED. THIS REPORT IS BEING GENERATED AT THE DIRECTION OF FDA BECAUSE OF SUPPLEMENTAL REPORTS RECEIVED, WITH THIS MANUFACTURER REPORT NUMBER, AND NO RECORD OF AN INITIAL REPORT. THIS MANUFACTURER REPORT IS SUBMITTED TO CORRECT A PREVIOUS REPORT, 3010157426-2014-00041, THAT MISTAKENLY IDENTIFIED THE MANUFACTURER REPORT NUMBER; AT FDA FORM 3500A MISTAKENLY IDENTIFIED THE MANUFACTURER BY LISTING THE MANUFACTURE¿S AFFILIATE ENTITY.

Description of Event or Problem · 1

SPACELABS RECEIVED A REPORT THAT AN ARKON ANESTHESIA MACHINE SPONTANEOUSLY WENT INTO A FAILED STATE DURING PATIENT USE ON (B)(6) 2015. NO ONE WAS INJURED AS A RESULT OF THIS EVENT.