G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2015-90042
- Event Type
- Malfunction
- Date Received
- September 18, 2015
- Date of Event
- August 27, 2015
- Report Date
- August 27, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- UDI-DI
- 00386270000019
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RECEIVER DATA LOG WAS REVIEWED ON 09/14/2015. THE COMPLAINT OF INACCURATE CGM VALUES WAS CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
(B)(4).
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT CONTINUOUS GLUCOSE MONITORING (CGM) INACCURACIES COMPARED TO BLOOD GLUCOSE (BG) METER THAT OCCURRED ON (B)(6) 2015. SENSOR WAS INSERTED ON (B)(6) 2015. AT THE TIME OF CONTACT, NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT RECEIVER WAS NOT RETURNED TO DEXCOM. THE TRANSMITTER (PART NUMBER STT-GL-003/SERIAL NUMBER (B)(4)/LOT NUMBER 5201333), BEING USED WITH THE COMPLAINT RECEIVER, WAS RETURNED ON (B)(4) 2015. THE RETURNED TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION. THE REPORTED EVENT OF INACCURACIES WAS NOT CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 618820 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | 9500-27 | ASKU | 00386270000019 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |