COULTER TQ-PREP WORKSTATION
Report
- Report Number
- 1061932-2015-01480
- Event Type
- Malfunction
- Date Received
- September 18, 2015
- Date of Event
- September 2, 2015
- Report Date
- September 2, 2015
- Manufacturer
- BECKMAN COULTER
- Product Code
- PER
- PMA / PMN Number
- K130253
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE CUSTOMER OBSERVED THAT THE SAMPLE TUBE DID NOT HAVE THE CORRECT AMOUNT OF REAGENT WHEN THE SAMPLE WAS PREP ON THE TQ PREP INSTRUMENT. THE SAMPLE LINE HAD ASPIRATED AIR SINCE THE REAGENT BOTTLES WERE EMPTY. THERE WAS A FAILURE OF THE SENSORS. NO ERROR MESSAGE ALERTED THE CUSTOMER THAT THE REAGENT BOTTLES WERE EMPTY SINCE THE SENSORS HAD FAILED. THE FSE REPLACED BOTH REAGENT SENSORS TO RESOLVE THE ISSUE.(B)(4).
THE FSE (FIELD SERVICE ENGINEER) REPORTED THAT THE REAGENT SENSORS (REAGENT A AND REAGENT B SENSORS) IN A COULTER TQ-PREP WORKSTATION FAILED AND DID NOT GENERATE THE LOW REAGENTS MESSAGES. THE CUSTOMER PERFORMS THE DAILY INSTRUMENT CHECK AS PER INSTRUCTIONS FOR USE AND THE CUSTOMER VISUALLY CONFIRMS THE TUBES AFTER PREPARATION, WHICH ALERTED THE CUSTOMER TO THE ISSUE. ERRONEOUS RESULTS WERE NOT GENERATED AND THERE WAS NO CHANGE OR EFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 619031 | COULTER TQ-PREP WORKSTATION | AUTOMATED PIPETTING, DILUTING AND SPECIMEN PROCESSING WORKSTATIONS FOR FLOW CYTO | PER | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |