FDA Adverse Event
Other
Summary report: N
VISUALPACS
MDR report key: 508665
·
Received January 9, 2004
Report
- Report Number
- 2135350-2004-00001
- Event Type
- Other
- Date Received
- January 9, 2004
- Date of Event
- December 10, 2003
- Report Date
- January 8, 2004
- Manufacturer
- EMAGEON UV, INC.
- Product Code
- LLZ
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
DURING A ROUTINE DEMONSTRATION AT A CUSTOMER SITE, IT WAS FOUND THAT DATA SETS THAT ARE TRANSFERRED FROM THE DICOM SERVER AS UN-CALIBRATED WITH DIFFERENT FIELDS OF VIEW FROM SLICE TO SLICE, CANNOT BE CALIBRATED CORRECTLY. THE PROBLEM IS EXHIBITED IN THE FOLLOWING WAY: IF AN IMAGE IS MANUALLY CALIBRATED USING ULTRAVISUAL TOOLS, THE OTHER SLICES (THAT HAVE DIFFERENT FIELDS OF VIEW) WILL INHERIT THE CALIBRATION OF THE ORIGINAL IMAGE. THE RESULT IS THAT IMAGES OTHER THAN THE INITIAL IMAGE CALIBRATED, WILL HAVE AN INCORRECT CALIBRATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VISUALPACS | MEDICAL IMAGE MANAGEMENT DEVICE | LLZ | EMAGEON UV, INC. | 2.2.1M | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |