FDA Adverse Event Malfunction Summary report: N

SPACELABS ULTRAVIEW SL COMMAND MODULE

MDR report key: 5086396 · Received September 17, 2015

Report

Report Number
3010157426-2015-00172
Event Type
Malfunction
Date Received
September 17, 2015
Date of Event
August 19, 2015
Report Date
November 16, 2015
Manufacturer
SPACELABS HEALTHCARE INC.
Product Code
DSI
PMA / PMN Number
K103142
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RECEIVED AT SPACELABS¿ EQUIPMENT SERVICE CENTER FOR REPAIR. THE REPORTED PROBLEM WAS VERIFIED. PART NUMBERS 333-0919-01 REAR MODULE PANEL, 670-0982-04 PCBA CPU WITH MPC 860 AND SDRA WERE USED TO REPLACE THE NON-FUNCTIONING ECG COMPONENTS. THE REPAIRED UNIT PASSED ALL FUNCTIONAL TESTS AND WAS RETURNED TO THE CUSTOMER. LACK OF A CONSISTENT ECG TRACE WAS THE WARNING SIGN THAT THE CUSTOMER NOTICED AND THAT PROMPTED THE ACTION OF REMOVING THE PRODUCT FROM USE. THIS REPORT IS COMPLETE AND THIS PARTICULAR ISSUE IS CONSIDERED CLOSED. SPACELABS MONITORS ITS COMPLAINTS FOR SIMILAR PRODUCT ISSUES TO DETERMINE NEED FOR ADDITIONAL INVESTIGATION.

Additional Manufacturer Narrative · 1

SPACELABS HAS LAUNCHED AN INVESTIGATION INTO THIS ISSUE AND WILL FILE A SUPPLEMENTAL REPORT ONCE THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

SPACELABS RECEIVED COMMAND MODULE MODEL 91496 FOR SERVICE REPAIR WITH CUSTOMER COMPLAINT OF "ECG KEEPS GOING ON AND OFF". THE CUSTOMER REMOVED THE PRODUCT FROM SERVICE ON (B)(6) 2015. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
616840 SPACELABS ULTRAVIEW SL COMMAND MODULE ULTRAVIEW SL MULTIPARAMETER MODULE DSI SPACELABS HEALTHCARE INC. 91496

Patients

Seq Age Sex Outcome Treatment
1