FDA Adverse Event Malfunction Summary report: N

CLINDAMYCIN CM 256 WW S30

MDR report key: 5085242 · Received September 17, 2015

Report

Report Number
3002769706-2015-00090
Event Type
Malfunction
Date Received
September 17, 2015
Report Date
July 2, 2015
Manufacturer
BIOMERIEUX, SA
Product Code
JWY
PMA / PMN Number
K913459/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE NOT RETURNED TO MANUFACTURER.

Additional Manufacturer Narrative · 1

BIOMÉRIEUX INTERNAL INVESTIGATION WAS CONDUCTED. THE RESULTS OF THE INVESTIGATION INDICATE RESULTS AT THE LIMIT OF THE COMPATIBILITY WITH ETEST® CLINDAMYCIN CM 256 (WHATEVER THE LOT TESTED) WHEN USED IN CONJUNCTION WITH MUELLER HINTON E AGAR. THIS ALLOWS A POTENTIAL FOR THE UNDERESTIMATION OF THE MIC FOR QC STRAINS 29213 WHICH CAN GIVE SOME VALUES OUT OF RANGE. BASED ON THE INVESTIGATION RESULTS, WHEN USING ETEST® CM 256 TEST KIT IN CONJUNCTION WITH MUELLER HINTON E AGAR, THERE IS A POTENTIAL FOR LOW MIC FOR STAPHYLOCOCCUS STRAINS. INTERMEDIATE (I) STRAINS COULD GIVE A FALSE SUSCEPTIBLE RESULT (S) OR RESISTANT (R) STRAINS COULD GIVE A FALSE INTERMEDIATE RESULT (I). FIELD CORRECTIVE ACTION (FCA-2744) AND CORRESPONDING PRODUCT CORRECTION NOTICE (CUSTOMER LETTER) WERE ISSUED TO THE FIELD 14-DEC-2015.

Description of Event or Problem · 1

A CUSTOMER IN (B)(6) CONTACTED BIOMERIEUX TO REPORT LOW RESULTS FOR ENTEROCOCCUS ATCC 29213 TM WITH ETEST CLINDAMYCIN CM 256 TEST KIT (REF 412315, LOT 1003867680). RETESTING PROVIDED THE SAME RESULT USING BOTH BIOMERIEUX AND OXOID ATCC 29213 TM. THE SAME ETEST CLINDAMYCIN CM 256 PRODUCT IS MARKETED IN THE UNITED STATES UNDER REFERENCE 412314. THERE IS NO INDICATION OR REPORT FROM THE HOSPITAL OR TREATING PHYSICIAN TO BIOMERIEUX THAT THE DISCREPANT RESULT LED TO ANY ADVERSE EVENT RELATED TO A PATIENT'S STATE OF HEALTH. CULTURE SUBMITTALS HAVE BEEN REQUESTED BY BIOMERIEUX FOR INTERNAL INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
615517 CLINDAMYCIN CM 256 WW S30 ETEST CLINDAMYCIN 256 JWY BIOMERIEUX, SA 1003867680

Patients

Seq Age Sex Outcome Treatment
1