FDA Adverse Event Malfunction Summary report: N

FAST-FIX 360 CURVED NDL DELIVERY SYS

MDR report key: 5084725 · Received September 17, 2015

Report

Report Number
1219602-2015-01012
Event Type
Malfunction
Date Received
September 17, 2015
Date of Event
August 27, 2015
Report Date
August 28, 2015
Manufacturer
SMITH & NEPHEW, INC.
Product Code
GAT
PMA / PMN Number
K092508
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE HAS BEEN RECEIVED, EVALUATION HAS NOT YET BEGUN. (B)(4).

Additional Manufacturer Narrative · 1

THREE USED FAST-FIX 360 CURVED NEEDLE DELIVERY SYSTEMS WERE RETURNED FOR EVALUATION. DEVICES WERE RETURNED IN THE SAME PACKAGING MAKING LOT IDENTIFICATION PROHIBITIVE. VISUAL ASSESSMENT OF EACH DEVICE SHOWED NO TS OR SUTURE REMAINS ON THE NEEDLES. ONE SUTURE/T CONSTRUCT WAS RETURNED. EXAMINATION OF THE CONSTRUCT SHOWED BOTH TS ARE BENT AND T1 IS MISSING ITS DISTAL END. DIMENSIONAL INSPECTION IS PROHIBITIVE AS A RESULT OF THIS DAMAGE. THE ACTUATOR ON ALL THREE DEVICES IS IN THE POST T2 DEPLOYMENT POSITION INDICATING A COMPLETE DEPLOYMENT. DEVICES WERE FUNCTIONALLY TESTED FOR PROPER ACTUATOR ADVANCEMENT AND CYCLING, EACH DEVICE FUNCTIONED AS INTENDED. NO ROOT CAUSE RELATED TO THE MANUFACTURING PROCESS CAN BE ESTABLISHED. NO FURTHER INVESTIGATION IS WARRANTED AT THIS TIME. (B)(4).

Description of Event or Problem · 1

DURING A MENISCAL REPAIR OF A POSTERIOR HORN FULL THICKNESS TEAR IT WAS REPORTED THAT WHEN HE TIGHTENED DOWN THE SUTURE THE IMPLANT PULLED OUT OF THE MENISCUS. HE REMOVED THE FAILED IMPLANT AND PROCEEDED TO USE ANOTHER WHICH ALSO PULLED OUT THE SAME WAY WHILE TIGHTENING IT DOWN. ANOTHER DEVICE WAS USED AND HAD THE SAME ISSUE. THERE WAS A FIFTEEN MINUTE DELAY IN THE PROCEDURE. NOTHING WAS LEFT IN THE PATIENT AND ALL DEBRIS WAS REMOVED. THERE WERE TWO ANCHORS THAT HELD AND THE DOCTOR FELT COMFORTABLE WITH THE REPAIR. THERE WAS NO PATIENT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
617060 FAST-FIX 360 CURVED NDL DELIVERY SYS SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE GAT SMITH & NEPHEW, INC. 50497476

Patients

Seq Age Sex Outcome Treatment
1 17 YR