FAST-FIX 360 CURVED NDL DELIVERY SYS
Report
- Report Number
- 1219602-2015-01012
- Event Type
- Malfunction
- Date Received
- September 17, 2015
- Date of Event
- August 27, 2015
- Report Date
- August 28, 2015
- Manufacturer
- SMITH & NEPHEW, INC.
- Product Code
- GAT
- PMA / PMN Number
- K092508
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
DEVICE HAS BEEN RECEIVED, EVALUATION HAS NOT YET BEGUN. (B)(4).
THREE USED FAST-FIX 360 CURVED NEEDLE DELIVERY SYSTEMS WERE RETURNED FOR EVALUATION. DEVICES WERE RETURNED IN THE SAME PACKAGING MAKING LOT IDENTIFICATION PROHIBITIVE. VISUAL ASSESSMENT OF EACH DEVICE SHOWED NO TS OR SUTURE REMAINS ON THE NEEDLES. ONE SUTURE/T CONSTRUCT WAS RETURNED. EXAMINATION OF THE CONSTRUCT SHOWED BOTH TS ARE BENT AND T1 IS MISSING ITS DISTAL END. DIMENSIONAL INSPECTION IS PROHIBITIVE AS A RESULT OF THIS DAMAGE. THE ACTUATOR ON ALL THREE DEVICES IS IN THE POST T2 DEPLOYMENT POSITION INDICATING A COMPLETE DEPLOYMENT. DEVICES WERE FUNCTIONALLY TESTED FOR PROPER ACTUATOR ADVANCEMENT AND CYCLING, EACH DEVICE FUNCTIONED AS INTENDED. NO ROOT CAUSE RELATED TO THE MANUFACTURING PROCESS CAN BE ESTABLISHED. NO FURTHER INVESTIGATION IS WARRANTED AT THIS TIME. (B)(4).
DURING A MENISCAL REPAIR OF A POSTERIOR HORN FULL THICKNESS TEAR IT WAS REPORTED THAT WHEN HE TIGHTENED DOWN THE SUTURE THE IMPLANT PULLED OUT OF THE MENISCUS. HE REMOVED THE FAILED IMPLANT AND PROCEEDED TO USE ANOTHER WHICH ALSO PULLED OUT THE SAME WAY WHILE TIGHTENING IT DOWN. ANOTHER DEVICE WAS USED AND HAD THE SAME ISSUE. THERE WAS A FIFTEEN MINUTE DELAY IN THE PROCEDURE. NOTHING WAS LEFT IN THE PATIENT AND ALL DEBRIS WAS REMOVED. THERE WERE TWO ANCHORS THAT HELD AND THE DOCTOR FELT COMFORTABLE WITH THE REPAIR. THERE WAS NO PATIENT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 617060 | FAST-FIX 360 CURVED NDL DELIVERY SYS | SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE | GAT | SMITH & NEPHEW, INC. | 50497476 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |