PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2015-44261
- Event Type
- Malfunction
- Date Received
- September 16, 2015
- Date of Event
- August 25, 2015
- Report Date
- August 25, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
THE INSULIN PUMP WAS RECEIVED STUCK IN THE MOTOR ERROR ALARM LOOP DURING BOLUS DELIVERY. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED; THE MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING OUR TESTING. UNABLE TO VERIFY E70 ALARM IN THE ALARM HISTORY DUE TO HISTORY FILE OVER WRITTEN. THE INSULIN PUMP PASSED REWIND, BASIC OCCLUSION, PRIME/A33 AND DISPLACEMENT TESTS. THE INSULIN PUMP HAS CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHED LCD WINDOW AND CRACKED BATTERY TUBE THREADS.
THE CUSTOMER CALLED IN TO REPORT THAT AFTER CHANGING THE BATTERY SHE RECEIVED A MOTOR ERROR ALARM AND A MOTOR POSITION ENCODER ERROR. THE CUSTOMER DID NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ISSUE. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED, AND WAS INFORMED TO RETURN THE PRODUCT FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 612200 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723RNAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR |