FDA Adverse Event
Malfunction
Summary report: N
PANORAMA CENTRAL STATION WITH TELEMETRY
MDR report key: 5081203
·
Received September 16, 2015
Report
- Report Number
- 3007222337-2015-00272
- Event Type
- Malfunction
- Date Received
- September 16, 2015
- Date of Event
- August 28, 2015
- Report Date
- August 28, 2015
- Manufacturer
- SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO. LTD.
- Product Code
- MHX
- PMA / PMN Number
- K142601
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
COMPANY REPRESENTATIVE EVALUATED THE SYSTEM. CORRECTIONS INCLUDED REPLACEMENT OF THE SYSTEM'S HARD DRIVE, CLEARING THE SYSTEM'S ERROR LOGS, AND REBOOTING. COMMUNICATION WAS REESTABLISHED. SYSTEM WAS SAFETY TESTED TO FACTORY'S SPECIFICATIONS. (B)(4) IS SUBMITTING THIS REPORT ON BEHALF OF SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. ((B)(4)).
Description of Event or Problem · 1
CUSTOMER REPORTED THE PANORAMA CENTRAL STATION WITH TELEMETRY'S SERVER SHUT DOWN, WHICH MAY HAVE AFFECTED TELEMETRY MONITORING. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 613456 | PANORAMA CENTRAL STATION WITH TELEMETRY | CENTRAL MONITORING STATION | MHX | SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO. LTD. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|