HOMECHOICE PRO
Report
- Report Number
- 1416980-2015-36184
- Event Type
- Malfunction
- Date Received
- September 16, 2015
- Date of Event
- August 22, 2015
- Report Date
- October 1, 2015
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RECEIVED AND THE EVALUATION IS COMPLETE. DURING THE EVENT HISTORY LOG REVIEW, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. A SERVICE HISTORY REVIEW REVEALED NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE EVENT. AN INTERNAL AND EXTERNAL VISUAL INSPECTION WAS PERFORMED. THE HOMECHOICE DEVICE RECEIVED A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE OF THE IIPV EVENT WAS DETERMINED TO BE A USE ERROR FURTHER SPECIFIED AS THE TOTAL TIDAL ULTRA FILTRATION REMOVAL WAS SET TOO LOW. THE HOMECHOICE APD SYSTEMS TRAINER¿S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). THE DEVICE HAS BEEN RECEIVED AND THE EVALUATION IS IN PROGRESS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2015 AT 21:26:42. DURING NIGHT DRAIN CYCLE FOUR, THE PATIENT'S ULTRAFILTRATION READING WAS 1380ML, INDICATING THE HOME PATIENT DRAINED 1275ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2100ML. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 612773 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |