STATSPIN SSMP CENTRIFUGE
Report
- Report Number
- 2023446-2015-00234
- Event Type
- Malfunction
- Date Received
- September 16, 2015
- Date of Event
- August 26, 2015
- Report Date
- August 26, 2015
- Manufacturer
- IRIS INTERNATIONAL
- Product Code
- JQC
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
BECKMAN PROVIDED A NEW BOARD AND MOTOR FOR THE CUSTOMER TO REPAIR THE UNIT. THE SYSTEM WAS NOT RETURNED TO MANUFACTURER FOR INSPECTION OR REPAIR. CUSTOMER REPAIRED THE UNIT ON SITE WITH PARTS SENT. THE ISSUE WAS RESOLVED WITH THEIR REPAIR. THE BECKMAN COULTER (BEC) INTERNAL IDENTIFIER FOR THIS REPORT IS (B)(4). NOT RETURNED TO MANUFACTURER.
THE CUSTOMER REPORTED AN BURNING SMELL FROM THEIR STATSPIN SSMP CENTRIFUGE DURING OPERATION. THE CUSTOMER COULD NOT CONFIRM IF THERE WERE SMOKE OR FIRES, BUT STATES IT WAS REPORTED THERE WAS AN ELECTRICAL BURNING ODOR. THERE WERE NO REPORTS OF DELAY TO SAMPLE PROCESSING. THERE WERE NO REPORTS OF INJURIES, NO DIRECT EXPOSURE AND NO MEDICAL ATTENTION WAS SOUGHT. THE CUSTOMER INDICATED THEY HAD REPLACED THE MOTOR ON THE UNIT JUST BEFORE THE EVENT. THE CUSTOMER INDICATED THEY ALSO SAW A WRENCH LIGHT ROTOR ERROR. THE CUSTOMER STARTED THE CYCLE AGAIN AND SMELLED THE BURNING SMELL. THE CUSTOMER OPENED THE UNIT AND OBSERVED THE PRINTED CIRCUIT BOARD (PCB) WAS BURNED AND ALSO OBSERVED THE BOARD WAS CHARRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 612668 | STATSPIN SSMP CENTRIFUGE | STATSPIN SSMP CENTRIFUGE | JQC | IRIS INTERNATIONAL | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |