FDA Adverse Event Malfunction Summary report: N

STATSPIN SSMP CENTRIFUGE

MDR report key: 5080627 · Received September 16, 2015

Report

Report Number
2023446-2015-00234
Event Type
Malfunction
Date Received
September 16, 2015
Date of Event
August 26, 2015
Report Date
August 26, 2015
Manufacturer
IRIS INTERNATIONAL
Product Code
JQC
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

BECKMAN PROVIDED A NEW BOARD AND MOTOR FOR THE CUSTOMER TO REPAIR THE UNIT. THE SYSTEM WAS NOT RETURNED TO MANUFACTURER FOR INSPECTION OR REPAIR. CUSTOMER REPAIRED THE UNIT ON SITE WITH PARTS SENT. THE ISSUE WAS RESOLVED WITH THEIR REPAIR. THE BECKMAN COULTER (BEC) INTERNAL IDENTIFIER FOR THIS REPORT IS (B)(4). NOT RETURNED TO MANUFACTURER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED AN BURNING SMELL FROM THEIR STATSPIN SSMP CENTRIFUGE DURING OPERATION. THE CUSTOMER COULD NOT CONFIRM IF THERE WERE SMOKE OR FIRES, BUT STATES IT WAS REPORTED THERE WAS AN ELECTRICAL BURNING ODOR. THERE WERE NO REPORTS OF DELAY TO SAMPLE PROCESSING. THERE WERE NO REPORTS OF INJURIES, NO DIRECT EXPOSURE AND NO MEDICAL ATTENTION WAS SOUGHT. THE CUSTOMER INDICATED THEY HAD REPLACED THE MOTOR ON THE UNIT JUST BEFORE THE EVENT. THE CUSTOMER INDICATED THEY ALSO SAW A WRENCH LIGHT ROTOR ERROR. THE CUSTOMER STARTED THE CYCLE AGAIN AND SMELLED THE BURNING SMELL. THE CUSTOMER OPENED THE UNIT AND OBSERVED THE PRINTED CIRCUIT BOARD (PCB) WAS BURNED AND ALSO OBSERVED THE BOARD WAS CHARRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
612668 STATSPIN SSMP CENTRIFUGE STATSPIN SSMP CENTRIFUGE JQC IRIS INTERNATIONAL NA NA