FDA Adverse Event Malfunction Summary report: N

NO FILAFORM INSERTER ENDOSCOPIC CAP STONE EXTRACTOR

MDR report key: 5079 · Received June 11, 1993

Report

Report Number
5079
Event Type
Malfunction
Date Received
June 11, 1993
Date of Event
April 21, 1993
Report Date
April 28, 1993
Manufacturer
COOK, INC.
Product Code
FFL
Product Problem
Yes
Report Source
User Facility report

Narratives

Description of Event or Problem · 1

APRIL 21, 1993, APPROX. 1600 HOURS, LAPAROSCOPIC CHOLECYSTECTOMY UNDERWAY WHEN A COMMON DUCT STONE IMPACTED IN THE AMPULLA OF VATER. BASKET STONE EXTRACTOR PASSED THROUGH CHOLEDOCHOSCOPE AND STONE WAS GRASPED. UNABLE TO FREE BASKET AS THERE APPEARED TO BE TRACTION ON THE CATHETER. NUMEROUS ATTEMPTS TO OPEN AND CLOSE AND RELEASE THE STONE WERE CARRIED OUT BUT STILL UNABLE TO GET RELEASE DEVICE TO WORK PROPERLY. GENTLE TRACTION UNSUCCESSFUL. MANIPULATIOLN UNSUCCESSFUL. CHOLEDOCHOSCOPE WAS REMOVED BY TAKING THE CLOSING AND OPENING DEVICE OFF THE END OF THE BASKET DEVICE. BASKET STILL WOULD NOT RELEASE. ANOTHER PULL ON THE BASKET RESULTED IN A SUDDEN RELEASE AND BASKET DEVICE WAS RETRACTED BUT THE ACTUAL WIRE BASKET PORTION WAS NOT VISIBLE. CHOLANGIOGRAM REVEALED BASKET IMPACTED IN THE AMPULLA AND STONE IN QUESTION ASSUMED TO HAVE BEEN PUSHED FORWARD. PROCEDURE COMPLETED (CHOLECYSTECTOMY). GASTROENTEROLOGIST IMMEDIATELY PERFORMED GASTRODUODENOSCOPY. AT THE SECOND PORTION OF THE DUODENUM, AT THE MAJOR AMPULLA, THE BASKET WAS PROTRUDING. IT WAS REMEOVED WITH BIOPSY FORCEPS VIA THE ENDOSCOPY THROUGH THE MOUTH, WITHOUT TRAUMA TO THE PATIENT. NO KNOWN PATIENT INJURY EXCEPT BEING SUBJECT TO A SECOND UNPLANNED PROCEDURE DUE TO WHAT SURGEON DESCRIBES AS A FRACTURE OF THE BASKETDEVICE LABELED FOR SINGLE USE. PATIENT MEDICAL STATUS PRIOR TO EVENT: SATISFACTORY CONDITION. THERE WAS NOT MULTIPLE PATIENT INVOLVEMENT.INVALID DATA - ON DEVICE SERVICE/MAINTENANCE. NO DATA - REGARDING DATE LAST SERVICED. SERVICE PROVIDED BY: INVALID DATA. INVALID DATA - SERVICE RECORDS AVAILABILITY. NO IMMINENT HAZARD TO PUBLIC HEALTH CLAIMED. DEVICE USED AS LABELED/INTENDED.DEVICE WAS EVALUATED AFTER THE EVENT. METHOD OF EVALUATION: ACTUAL DEVICE INVOLVED IN INCIDENT WAS EVALUATED, VISUAL EXAMINATION. RESULTS OF EVALUATION: TELEMETRY FAILURE, NONE OR UNKNOWN, NONE OR UNKNOWN, NONE OR UNKNOWN. CONCLUSION: NONE OR UNKNOWN. CERTAINTY OF DEVICE AS CAUSE OF OR CONTRIBUTOR TO EVENT: YES. CORRECTIVE ACTIONS: DEVICE PERMANENTLY REMOVED FROM SERVICE. THE DEVICE WAS NOT DESTROYED/DISPOSED OF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NO FILAFORM INSERTER ENDOSCOPIC CAP STONE EXTRACTOR 4 WIRE, 115 CM, 3FR FLAT WIRE STONE EXTRACTOR FFL COOK, INC. 0633515

Patients

Seq Age Sex Outcome Treatment
1 45 YR Other