FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 5078664 · Received September 15, 2015

Report

Report Number
2032227-2015-43734
Event Type
Injury
Date Received
September 15, 2015
Date of Event
August 24, 2015
Report Date
August 24, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
LA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH CRACKED AND BLEEDING LCD GLASS, DAMAGE ON ELECTRONIC ASSEMBLY ON MOTHER BOARD AND INTERFACE BOARD, MISSING REMOTE FREQUENCY BOARD AND BROKEN OFF THE END CAP. IT WAS UNABLE TO PERFORM FUNCTIONAL TESTINGS INCLUDING DUE TO DAMAGED ELECTRONIC ASSEMBLY. (B)(4)

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT HE WAS HOSPITALIZED FROM (B)(6) 2015 WITH HIGH BLOOD GLUCOSE OF 474 MG/DL. THE CUSTOMER WAS NOT WEARING THE INSULIN PUMP AT THE TIME OF THE HOSPITALIZATION. HE HAD BEEN OFF FOR 2 OR 3 HOURS. THE CUSTOMER EXPLAINED THAT HE HAD GONE MUD RIDING AND FOUND THE INSULIN PUMP IN THE MUD; IT WAS SHATTERED AND THE MOTOR CAME OUT. CUSTOMER STATED THAT IT MIGHT HAVE BEEN RUN OVER. BLOOD GLUCOSE VALUE AT THE TIME WAS 115 MG/DL. THE INSULIN PUMP WAS REPLACED AND RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
608704 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAP

Patients

Seq Age Sex Outcome Treatment
1 22 YR Hospitalization