BREAS VIVO 60
Report
- Report Number
- 9617566-2015-00135
- Event Type
- Malfunction
- Date Received
- September 15, 2015
- Date of Event
- January 1, 2014
- Report Date
- September 14, 2015
- Manufacturer
- BREAS MEDICAL AB
- Product Code
- NOU
- PMA / PMN Number
- K123144
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
UNDER EUROPEAN LAW, PATIENT INFORMATION IS CONSIDERED CONFIDENTIAL AND WILL NOT BE RELEASED BY THE HOSPITAL. THE VIVO 60 IS NOT MARKETED IN THE USA, BUT THIS REPORT IS BEING SUBMITTED TO THE FDA SINCE THE VIVO 50 IN THE US IS A SIMILAR MODEL. VIVO 60 DOES NOT HAVE A PRODUCT CODE LISTED SO THE VIVO 50 PRODUCT CODE NOU IS USED. THE VIVO 50 510(K) NUMBER IS K123144. THIS MEDWATCH REPORT IS A RESULT OF A RETROSPECTIVE INVESTIGATION OF CUSTOMER COMPLAINTS AND HAS BEEN DEEMED REPORTABLE DUE TO INSUFFICIENT PATIENT INFORMATION. BASED UPON THE INFORMATION THAT BREAS PRESENTLY POSSESSES, IT IS UNABLE TO CONFIRM WHETHER THE EVENT IS REPORTABLE UNDER THE STANDARD SET FORTH IN THE APPLICABLE FDA REGULATIONS AND GUIDANCE. ACCORDINGLY, IN AN ABUNDANCE OF CAUTION, BREAS IS FILING A MEDWATCH REPORT FOR THIS EVENT. BEING PART OF A RETROSPECTIVE ANALYSIS, THIS REPORT IS SUBMITTED LATER THAN 30 DAYS AFTER THE INITIAL DATE ON WHICH BREAS BECAME AWARE OF THE EVENT, AS HAS PREVIOUSLY BEEN DISCUSSED WITH THE FDA.
BREAS (B)(4) REPORTED A VIVO 60 DEVICE WITH A BATTERY ISSUE. THE INTERNAL BATTERY ICON HAD A RED CROSS LIKE IF IT'S OUT OF ORDER. WHEN WE HAVE RECONNECTED THE MAINS POWER, THE INTERNAL BATTERY ICON CAME BACK IN GREEN AND THE CHARGE WAS RUNNING. THE VIVO 60 IS NOT MARKETED IN THE USA, BUT THIS REPORT IS BEING SUBMITTED TO THE FDA SINCE THE VIVO 50 IN THE US IS A SIMILAR MODEL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 608986 | BREAS VIVO 60 | VIVO 60 CONTINUOUS VENTILATOR | NOU | BREAS MEDICAL AB | VIVO 60 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |