FDA Adverse Event
Malfunction
Summary report: N
SPIRIT 3000 DENTAL CHAIR
MDR report key: 5075914
·
Received September 14, 2015
Report
- Report Number
- 1017522-2015-00020
- Event Type
- Malfunction
- Date Received
- September 14, 2015
- Date of Event
- August 17, 2015
- Report Date
- August 17, 2015
- Manufacturer
- PELTON & CRANE
- Product Code
- NRU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- DENTAL HYGIENIST
Narratives
Additional Manufacturer Narrative · 1
UPON VISUAL INSPECTION BY THE AUTHORIZED (B)(4) DISTRIBUTOR, IT WAS IDENTIFIED THAT THE FOUR MOUNTING BOLTS THAT ATTACHES THE REAR CUSPIDOR BRACKET TO THE DENTAL CHAIR WERE LOOSE WHICH CAUSED THE TOP TWO MOUNTING BOLTS TO BREAK. THE DISTRIBUTOR HAS ALREADY DISPOSED OF THE BOLTS THEREFORE (B)(4) CANNOT COMPLETE THE EVALUATION. THE DENTAL CHAIR IS OVER 10 1/2 YEARS OLD AND IS PAST THE EXPECTED LIFE OF THE DEVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT A DENTAL HYGIENIST WAS PERFORMING ROUTINE MEDICAL TREATMENT TO A PATIENT WHEN THE REAR MOUNTED CUSPIDOR FELL OFF THE (B)(4) SP30 DENTAL CHAIR. THERE WERE NO INJURIES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 605632 | SPIRIT 3000 DENTAL CHAIR | DENTAL CHAIR | NRU | PELTON & CRANE | SP30 |