FDA Adverse Event Malfunction Summary report: N

SCR BIORCI 7X25MM (8MM HEAD) STRL BX 1

MDR report key: 5073082 · Received September 12, 2015

Report

Report Number
1219602-2015-01002
Event Type
Malfunction
Date Received
September 12, 2015
Date of Event
May 23, 2014
Report Date
February 11, 2016
Manufacturer
SMITH & NEPHEW, INC.
Product Code
HWC
PMA / PMN Number
K992396
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

ONE SCR, BIORCI 7X25MM (8MM HEAD) WAS RETURNED FOR EVALUATION OF THE REPORTED COMPLAINT MODE, ¿BROKEN SCREW.¿ VISUAL INSPECTION CONFIRMED THE REPORTED COMPLAINT; THE SCREW WAS BROKEN INTO SEVERAL PIECES. BASED ON THE DETAILS OF THE COMPLAINT, THE LIKELY CAUSE OF THE SCREW BREAKING IS DUE A LACK OF HOLE PREPARATION PRIOR TO SCREW INSERTION. PER IFU 1060635, ¿THE STARTER MUST BE UTILIZED WITH THE BIORCI SCREWS TO MINIMIZE SCREW BREAKAGE DURING INSERTION.¿ DUE TO THESE OBSERVATIONS NO ROOT CAUSE RELATED TO THE MANUFACTURING PROCESS CAN BE ESTABLISHED. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED WHICH CONFIRMED NO INCONSISTENCIES. THERE WERE NO INTERNAL PROCESSING ISSUES, WHICH COULD HAVE CONTRIBUTED TO THE NATURE OF THE COMPLAINT. A COMPLAINT HISTORY REVIEW HAS NOT IDENTIFIED ADDITIONAL COMPLAINTS FOR THIS LOT NUMBER ON FILE. NO FURTHER INVESTIGATION IS WARRANTED AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

INDIVIDUAL MDR SUBMISSION NOT REQUIRED. THIS EVENT WAS PREVIOUSLY FILED UNDER ASR (B)(4) ON JULY 31, 2014.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ART KNEE PROCEDURE USING SCR BIORCI 7X25MM (8MM HEAD) STRL BX 1 THE DISTAL END OF THE SCREW BROKE. DEBRIS WAS REMOVED FROM THE PATIENT. THERE WAS A PROCEDURAL DELAY OF 3.5 HOURS. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. THIS INCIDENT DID NOT RESULT IN ANY PATIENT INJURY OR COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
604410 SCR BIORCI 7X25MM (8MM HEAD) STRL BX 1 SCREW, FIXATION, BONE HWC SMITH & NEPHEW, INC. 50406116

Patients

Seq Age Sex Outcome Treatment
1