FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 5071631 · Received September 11, 2015

Report

Report Number
2531779-2015-32560
Event Type
Malfunction
Date Received
September 11, 2015
Report Date
September 9, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/09/2015 WITH THE FOLLOWING FINDINGS: INVESTIGATION REVEALED THAT THE KEYPAD WAS PEELING. THE CONTRAST BUTTON WAS UNRESPONSIVE AND THE UP ARROW KEYPAD BUTTON WAS INTERMITTENTLY UNRESPONSIVE. THE KEYPAD COVER WAS REMOVED AND THERE WAS EVIDENCE OF CONTAMINATION UNDER THE UP ARROW BUTTON CONTACT. ADDITIONALLY, THE CONTRAST BUTTON CONTACT WAS MISSING. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THERE WAS CONTAMINATION UNDER THE BUTTON CONTACTS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 09/09/2015.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
603424 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 24