FDA Adverse Event Malfunction Summary report: N

ACTIVA

MDR report key: 5070839 · Received September 11, 2015

Report

Report Number
3004209178-2015-17807
Event Type
Malfunction
Date Received
September 11, 2015
Report Date
April 28, 2016
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
MHY
PMA / PMN Number
P960009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE DESKTOP CHARGER ((B)(4)) FOUND THE CABLE ASSEMBLY TO HAVE A DAMAGED TIP.

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID 37761, SERIAL# (B)(4), PRODUCT TYPE: RECHARGER. PRODUCT ID 3389S-40, LOT# VA0AB21, IMPLANTED: (B)(6) 2013, PRODUCT TYPE: LEAD. PRODUCT ID 748251, SERIAL# (B)(4), IMPLANTED: (B)(6) 2007, PRODUCT TYPE: EXTENSION. PRODUCT ID 748251, SERIAL# (B)(4), IMPLANTED: (B)(6) 2007, PRODUCT TYPE: EXTENSION. PRODUCT ID 3387-40, LOT# V000525, IMPLANTED: (B)(6) 2006, PRODUCT TYPE: LEAD. (B)(4).

Description of Event or Problem · 1

INFORMATION WAS RECEIVED FROM THE CONSUMER THAT REPORTED THERE WAS A BROKEN CONNECTOR PIN. THE PATIENT HAD SHAKING, ISSUES WALKING, AND THEY COULD NOT FEED THEMSELVES. THE CHANGE IN THERAPY/SYMPTOMS WAS SUDDEN. THE SYMPTOMS WERE BECAUSE THE PATIENT'S IMPLANTABLE NEUROSTIMULATOR (INS) WAS DEPLETED AND HE NEEDED TO RECHARGE. THE LAST SUCCESSFUL RECHARGE WAS SOMETIME A WEEK PRIOR TO REPORT. ADDITIONAL INFORMATION RECEIVED FROM THE MANUFACTURER REPRESENTATIVE REPORTED THE BATTERIES IN THE DEEP BRAIN STIMULATOR (DBS) WERE NOT CHARGING CORRECTLY. THE PATIENT WAS IMPLANTED FOR ESSENTIAL TREMOR AND MOVEMENT DISORDERS. FURTHER FOLLOW-UP IS BEING CONDUCTED TO DETERMINE WHAT TROUBLESHOOTING OCCURRED AND WHAT ACTIONS/INTERVENTIONS WERE TAKEN TO RESOLVE THE ISSUE OF THE DBS BATTERIES NOT CHARGING CORRECTLY AND IF IT WAS ONLY DUE TO THE BROKEN CONNECTOR PIN IN THE DESKTOP CHARGER. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
603620 ACTIVA STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR MHY MEDTRONIC PUERTO RICO OPERATIONS CO. 37612

Patients

Seq Age Sex Outcome Treatment
1 00075 YR