FDA Adverse Event Malfunction Summary report: N

CARE ASSIST CA305

MDR report key: 5070704 · Received September 11, 2015

Report

Report Number
9615739-2015-00271
Event Type
Malfunction
Date Received
September 11, 2015
Date of Event
August 20, 2015
Report Date
August 20, 2015
Manufacturer
HILL-ROM FRANCE
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
HOSPITAL SERVICE TECHNICIAN

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND 4 BROKEN CASTER SUPPORTS, 2 BROKEN BRAKE BAR SPACERS AND 1 BROKEN CASTER. PER THE HILL-ROM SERVICE MANUAL THE BED SHOULD BE SUBJECT TO AN EFFECTIVE MAINTENANCE PROGRAM. AN ANNUAL SERVICE OF THE BED IS ADVISED IN ORDER TO MAINTAIN ITS CHARACTERISTICS AND PERFORMANCE. BRAKE CASTERS SHOULD BE CHECKED FOR CUTS, WEAR AND QUALITY OF TREAD, ETC. AND REPLACED WHEN NECESSARY. CHECK THE BRAKES TO SEE WHETHER THE BED MOVES WHEN THE BRAKE PEDALS ARE PRESSED AND REPAIR AS NECESSARY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2015. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE CASTER SUPPORTS, BRAKE BAR SPACERS AND CASTER TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES WERE NOT HOLDING. THE BED WAS LOCATED IN MEDSURG AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
603553 CARE ASSIST CA305 A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM FRANCE P1170C0000008

Patients

Seq Age Sex Outcome Treatment
1