NICOLET VEEG SYSTEM
Report
- Report Number
- 3010611950-2015-00018
- Event Type
- Injury
- Date Received
- September 10, 2015
- Date of Event
- October 24, 2014
- Report Date
- October 24, 2014
- Manufacturer
- NATUS NEUROLOGY, INCORPORATED
- Product Code
- GWQ
- PMA / PMN Number
- K964280
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THIS IS A RE-FILING OF REPORT 3008289288-2014-00005 WITH A NEW REPORT NUMBER. THE REPORT WAS EDITED ON (B)(4) 2015 TO REVISE THE FACILITY REGISTRATION NUMBER FROM 3008289288 TO 3010611950. THE CURRENT FACILITY REGISTRATION NUMBER CHANGED IN MAY 2014 WHEN NATUS NEUROLOGY INC. MOVED TO ITS CURRENT LOCATION AT (B)(4). PATIENT INFORMATION: PATIENT INFORMATION WAS NOT PROVIDED, PATIENT AGE AND WEIGHT ARE THEREFORE DEFAULTED TO "0". DEVICE EVALUATED BY MANUFACTURER: DEVICE WAS NOT RETURNED TO MANUFACTURER FOR EVALUATION. REGISTERED INTERNALLY AS A COMPLAINT (REFERENCE (B)(4)) FOR FURTHER INVESTIGATION BY MANUFACTURER. THE INTERNAL INVESTIGATION CONFIRMED THAT THE CART DESIGN HAD BEEN SUCCESSFULLY TIP TESTED BY THE MANUFACTURER AND THAT THE SYSTEM IN QUESTION MAY HAVE CAUGHT ON A THRESHOLD REQUIRING THAT AN UNUSUAL AMOUNT OF FORCE BE USED TO TRY AND MOVE IT WHICH INADVERTENTLY TIPPED THE CART.
A HOSPITAL TECHNICIAN WAS MOVING A NATUS NEUROLOGY NICONE VEEG MONITORING SYSTEM ON A ROLL-ABLE ERGOTRON CART INTO/THROUGH AN ELEVATOR. THE CART TIPPED OVER INJURING THE TECHNICIAN. CART LABELING REQUIRES THAT THE SYSTEM MUST NOT EXTEND UP OR OUT AND THAT CARE MUST BE USED WHEN PUSHING OR PULLING CARTS OVER DEPRESSED SURFACES. THERE WAS NO PATIENT INVOLVEMENT AT THE TIME THE INCIDENT OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 598490 | NICOLET VEEG SYSTEM | ELECTROCEPHLOGRAPH SYSTEM | GWQ | NATUS NEUROLOGY, INCORPORATED | 982A0506 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR | Other |