FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 5066524 · Received September 9, 2015

Report

Report Number
2531779-2015-32104
Event Type
Malfunction
Date Received
September 9, 2015
Report Date
August 19, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION 10/08/2015-PRODUCT ANALYSIS:THE DEVICE WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/23/2015 WITH THE FOLLOWING FINDINGS:THE COMPLAINT COULD NOT BE DUPLICATED WITH INVESTIGATION. REVIEW OF THE PUMP¿S BLACK BOX REVEALED RELATED ALARMS. THE PUMP SUCCESSFULLY COMPLETED A PRIME SEQUENCE AND 24-HOUR EXERCISE TEST WITHOUT ISSUE OR ALARM. THE PUMP¿S FORCE SENSOR CALIBRATION WAS FOUND TO BE WITHIN SPECIFICATION.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A FREQUENT PERSISTENT OCCLUSION ALARM ISSUE. IT WAS REPORTED THE PATIENT'S BLOOD GLUCOSE WAS BETWEEN 250-499 MG/DL WITHOUT SIGNS OR SYMPTOMS OF HYPERGLYCEMIA. THE REPORTED HEALTH EVENT DOES NOT QUALIFY AS A SERIOUS INJURY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE REMAINED UNRESOLVED AT THE TIME OF TROUBLE SHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
595226 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 21 YR