FDA Adverse Event Malfunction Summary report: N

ACCESSORIES FOR SPACE PUMPS

MDR report key: 5065147 · Received September 9, 2015

Report

Report Number
9610825-2015-00385
Event Type
Malfunction
Date Received
September 9, 2015
Date of Event
August 28, 2015
Report Date
August 28, 2015
Manufacturer
B. BRAUN MELSUNGEN AG
Product Code
MRZ
PMA / PMN Number
K083689
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ACTUAL SAMPLE INVOLVED HAS BEEN RECEIVED FOR EVALUATION AND THE INVESTIGATION IS ONGOING AT THIS TIME. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION RESULTS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). ONE SAMPLE WAS RECEIVED AND ALL AVAILABLE INFORMATION WAS FORWARDED TO THE MANUFACTURER B. BRAUN MELSUNGEN. THE OUTER LIVE PIN WAS BENT AND THERMAL DAMAGE WAS FOUND ON THE POWER CORD. THE PINS OF THE POWER CORD THAT WAS PLUGGED INTO THE WALL MAY HAVE BEEN IN CONTACT WITH A CONDUCTIVE FLUID; ALSO A MECHANICAL DAMAGE COULD NOT BE EXCLUDED. THE REASON FOR THE THERMAL DAMAGE WAS CAUSED BY AN ELECTROLYSIS UNLEASHED BY A FLUID INGRESS BETWEEN THE CONTACT OF THE WALL SOCKET AND THE EXPECTED WET CONDUCTIVE PINS. BECAUSE OF THE CAPILLARY EFFECT THE LIQUID HAS THE ABILITY TO GET TO THE CONDUCTIVE PARTS OF THE POWER SUPPLY. IT SHOULD BE NOTED THAT THE INSTRUCTION FOR USE (IFU) FOR THE SPACE PUMP STATE "PROTECT THE DEVICE AND THE POWER SUPPLY FROM MOISTURE" AND ISO SYMBOL TO PROTECT FROM MOISTURE IS DISPLAYED ON THE POWER SUPPLY.

Description of Event or Problem · 1

AS REPORTED BY THE USER FACILITY: THE CUSTOMER REPORTED "BROKEN PRONGS ON CHARGER" HOWEVER UPON RECEIPT OF THE SAMPLES ONE OF THE SAMPLES EXHIBITED THERMAL DAMAGE. NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
596998 ACCESSORIES FOR SPACE PUMPS US. POWER CORD INCL. EXTENSION MRZ B. BRAUN MELSUNGEN AG 8713112D

Patients

Seq Age Sex Outcome Treatment
1