ACCESSORIES FOR SPACE PUMPS
Report
- Report Number
- 9610825-2015-00385
- Event Type
- Malfunction
- Date Received
- September 9, 2015
- Date of Event
- August 28, 2015
- Report Date
- August 28, 2015
- Manufacturer
- B. BRAUN MELSUNGEN AG
- Product Code
- MRZ
- PMA / PMN Number
- K083689
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE ACTUAL SAMPLE INVOLVED HAS BEEN RECEIVED FOR EVALUATION AND THE INVESTIGATION IS ONGOING AT THIS TIME. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION RESULTS BECOME AVAILABLE.
(B)(4). ONE SAMPLE WAS RECEIVED AND ALL AVAILABLE INFORMATION WAS FORWARDED TO THE MANUFACTURER B. BRAUN MELSUNGEN. THE OUTER LIVE PIN WAS BENT AND THERMAL DAMAGE WAS FOUND ON THE POWER CORD. THE PINS OF THE POWER CORD THAT WAS PLUGGED INTO THE WALL MAY HAVE BEEN IN CONTACT WITH A CONDUCTIVE FLUID; ALSO A MECHANICAL DAMAGE COULD NOT BE EXCLUDED. THE REASON FOR THE THERMAL DAMAGE WAS CAUSED BY AN ELECTROLYSIS UNLEASHED BY A FLUID INGRESS BETWEEN THE CONTACT OF THE WALL SOCKET AND THE EXPECTED WET CONDUCTIVE PINS. BECAUSE OF THE CAPILLARY EFFECT THE LIQUID HAS THE ABILITY TO GET TO THE CONDUCTIVE PARTS OF THE POWER SUPPLY. IT SHOULD BE NOTED THAT THE INSTRUCTION FOR USE (IFU) FOR THE SPACE PUMP STATE "PROTECT THE DEVICE AND THE POWER SUPPLY FROM MOISTURE" AND ISO SYMBOL TO PROTECT FROM MOISTURE IS DISPLAYED ON THE POWER SUPPLY.
AS REPORTED BY THE USER FACILITY: THE CUSTOMER REPORTED "BROKEN PRONGS ON CHARGER" HOWEVER UPON RECEIPT OF THE SAMPLES ONE OF THE SAMPLES EXHIBITED THERMAL DAMAGE. NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 596998 | ACCESSORIES FOR SPACE PUMPS | US. POWER CORD INCL. EXTENSION | MRZ | B. BRAUN MELSUNGEN AG | 8713112D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |