FDA Adverse Event Malfunction Summary report: N

INRATIO2 PT/INR PROFESSIONAL TEST STRIP

MDR report key: 5063999 · Received September 9, 2015

Report

Report Number
2027969-2015-00660
Event Type
Malfunction
Date Received
September 9, 2015
Date of Event
August 13, 2015
Report Date
August 13, 2015
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
GJS
PMA / PMN Number
K110212
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION IS PENDING.

Additional Manufacturer Narrative · 1

INVESTIGATION/CONCLUSION: IT IS INDICATED THAT THE PRODUCT IS NOT RETURNING FOR EVALUATION. THEREFORE, A REVIEW OF IN-HOUSE TESTING WAS PERFORMED. IN-HOUSE TESTING ON STRIP LOT K370019 MEETS RELEASE CRITERIA. THE PRODUCT PERFORMED AS EXPECTED. A REVIEW OF THE MANUFACTURING RECORDS FOR LOT# K370019 DID NOT UNCOVER ANY RELEVANT NON-CONFORMANCES. THE LOT MEETS RELEASE SPECIFICATIONS. ROOT CAUSE COULD NOT BE DETERMINED FROM THE INFORMATION PROVIDED BY THE CUSTOMER. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.

Description of Event or Problem · 1

A PHYSICIAN'S OFFICE IN (B)(6) ALLEGED A VARIANCE BETWEEN THE INRATIO2 INR RESULT AND THE LABORATORY INR RESULT. RESULTS ARE AS FOLLOWS: INRATIO2 INR: 1.4, LABORATORY INR: 2.1. THERAPEUTIC RANGE: 2.0 - 3.0. THE TIME BETWEEN TESTING WAS A FEW MINUTES. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THERE WAS NO ADDITIONAL INFORMATION PROVIDED. (NOTE: THE INRATIO2 PRODUCT 99008G1 IS NOT AVAILABLE IN THE UNITED STATES; HOWEVER, THIS MDR FILING IS DUE TO THE A SAME OR SIMILAR DEVICE BEING AVAILABLE IN THE UNITED STATES.)

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
595043 INRATIO2 PT/INR PROFESSIONAL TEST STRIP PROTHROMBIN TIME TEST GJS ALERE SAN DIEGO, INC. 99008G1 K370019

Patients

Seq Age Sex Outcome Treatment
1