FDA Adverse Event
Malfunction
Summary report: N
POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI
MDR report key: 5063741
·
Received September 9, 2015
Report
- Report Number
- 1525712-2015-04475
- Event Type
- Malfunction
- Date Received
- September 9, 2015
- Report Date
- November 10, 2015
- Manufacturer
- INVACARE TAYLOR STREET
- Product Code
- ITI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
ADDITIONAL/UPDATED INFORMATION WAS ADDED TO REFLECT THE DEVICE BEING RETURNED TO THE MANUFACTURER FOR EVALUATION. THE MOTORS, CONTROLLER, AND JOYSTICK WERE RETURNED FOR EVALUATION, HOWEVER SUBSEQUENT TESTING COULD NOT VERIFY THE COMPLAINT. THE CONTROLLER AND JOYSTICK PASSED FUNCTIONAL TESTING, AND BOTH MOTORS OPERATED PROPERLY DURING THE BENCH TEST, RUNNING IN THE PROPER DIRECTIONS.
Additional Manufacturer Narrative · 1
SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A SUPPLEMENTAL RECORD WILL BE FILED.
Description of Event or Problem · 1
DEALER STATES WHEN THE JOYSTICK IS IN FORWARD IT GOES REVERSE, IN REVERSE IT GOES FORWARD, RIGHT GOES RIGHT, LEFT GOES REVERSE.
Description of Event or Problem · 1
DEALER STATES WHEN THE JOYSTICK IS IN FORWARD IT GOES REVERSE, IN REVERSE IT GOES FORWARD, RIGHT GOES RIGHT, LEFT GOES REVERSE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 594758 | POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI | WHEELCHAIR, POWERED | ITI | INVACARE TAYLOR STREET | 3GAR-CG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |