FDA Adverse Event Injury Summary report: N

LEAD EXTRACTION EVOLUTION SHORTIE RL CONTROLLED-ROTATION DILATOR SHEATH SET

MDR report key: 5063617 · Received September 9, 2015

Report

Report Number
1820334-2015-00573
Event Type
Injury
Date Received
September 9, 2015
Date of Event
August 13, 2015
Report Date
August 13, 2015
Manufacturer
COOK VASCULAR INC
Product Code
DRE
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PMA/510(K) # K061000. EVENT EVALUATION: NO PRODUCT WAS RETURNED TO ASSIST IN THE INVESTIGATION; HOWEVER, A REVIEW OF THE COMPLAINT HISTORY AND DEVICE HISTORY RECORD WAS CONDUCTED. THERE IS NO EVIDENCE TO SUGGEST THE PRODUCT WAS NOT MANUFACTURED TO SPECIFICATION. THERE IS NO INFORMATION TO SUGGEST THE DEVICE MALFUNCTIONED OR CONTAINED A NONCONFORMITY. WORK ORDERS WERE REVIEWED, AND NO EVIDENCE OF NONCONFORMITY WAS FOUND. COMPLAINTS WERE REVIEWED, AND NO OTHER COMPLAINTS HAVE BEEN FILED AGAINST DEVICES FROM THESE LOTS. BECAUSE THE DEVICES WERE NOT RETURNED, NO IMAGES WERE PROVIDED, AND FEW DETAILS OF THE PROCEDURE WERE PROVIDED, THE ROOT CAUSE OF THE COMPLAINT IS UNKNOWN. DUE TO THE NATURE OF LEAD EXTRACTION SURGERY, A DEVICE CAN FUNCTION NORMALLY AND STILL CAUSE A TEAR IN A BLOOD VESSEL. WE WILL CONTINUE TO MONITOR THIS DEVICE.

Additional Manufacturer Narrative · 1

(B)(4). THE EVENT IS CURRENTLY UNDER INVESTIGATION.

Description of Event or Problem · 1

DURING LEAD EXTRACTION OF A RV LEAD, THE SUBCLAVIAN VEIN WAS TORN. THE PATIENT WAS PUT ON BYPASS AND UNDERWENT OPEN CHEST SURGERY TO REPAIR DAMAGE TO THE SUBCLAVIAN VEIN AND SVC. NO ADDITIONAL INFORMATION WAS PROVIDED REGARDING PATIENT OUTCOME.

Description of Event or Problem · 1

DURING LEAD EXTRACTION OF A RV LEAD, THE SUBCLAVIAN VEIN WAS TORN. THE PATIENT WAS PUT ON BYPASS AND UNDERWENT OPEN CHEST SURGERY TO REPAIR DAMAGE TO THE SUBCLAVIAN VEIN AND SVC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
596500 LEAD EXTRACTION EVOLUTION SHORTIE RL CONTROLLED-ROTATION DILATOR SHEATH SET DRE DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION DRE COOK VASCULAR INC N/A

Patients

Seq Age Sex Outcome Treatment
1 84 YR