FDA Adverse Event Malfunction Summary report: N

UNICEL DXC 600 SYNCHRON SYSTEM

MDR report key: 5062091 · Received September 8, 2015

Report

Report Number
2050012-2015-00341
Event Type
Malfunction
Date Received
September 8, 2015
Date of Event
August 19, 2015
Report Date
August 19, 2015
Manufacturer
BECKMAN COULTER
Product Code
JJE
PMA / PMN Number
K103842
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT AND FOUND THE WASH VALVE ON THE BUBBLE GENERATOR ASSEMBLY WAS LEAKING. THE FSE REPLACED THE BUBBLE GENERATOR ASSEMBLY TO RESOLVE THE ISSUE, NO FURTHER LEAKING OR INSTRUMENT ERRORS WERE OBSERVED. THE INSTRUMENT SOFTWARE VERSION IS 5.4.

Description of Event or Problem · 1

THE CUSTOMER'S UNICEL DXC 600 SYNCHRON SYSTEM GENERATED ONE LOW ERRONEOUS CR-S (CREATININE) PATIENT RESULT. THE CR-S ERRONEOUS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY, HOWEVER THE RESULT WAS RERUN AND AMENDED WITH NO IMPACT TO PATIENT TREATMENT. THE CUSTOMER ALSO STATED THE INSTRUMENT GENERATED SUPPRESSED (NO RESULT VALUE) QC (QUALITY CONTROL) AND RECOVERY ERRORS FOR COCM (COCAINE METABOLITE), ALT (ALANINE AMINOTRANSFERASE), BUN (BLOOD UREA NITROGEN) AND FOUND FLUID UNDER THE REAGENT PROBE DRIP TRAY. THE CUSTOMER STATED THE FLUID WAS CONTAINED AND DESCRIBED THE VOLUME AS APPROXIMATELY LESS THAN 1 CC. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT AND NO INJURY OR EXPOSURE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
592574 UNICEL DXC 600 SYNCHRON SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1 77 YR