CUSA EXCEL 36KHZ CEM NOSECONE
Report
- Report Number
- 3006697299-2015-00139
- Event Type
- Malfunction
- Date Received
- September 8, 2015
- Date of Event
- July 10, 2015
- Report Date
- August 17, 2015
- Manufacturer
- INTEGRA LIFESCIENCES(IRELAND)
- Product Code
- LFL
- PMA / PMN Number
- K981262
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
INTEGRA COMPLETED ITS INTERNAL INVESTIGATION 09/23/2015. THE INVESTIGATION INCLUDED: METHOD: DHR REVIEW, REVIEW OF COMPLAINT MANAGEMENT DATABASE FOR SIMILAR COMPLAINTS. VISUAL INSPECTION. RESULTS: DEVICE HISTORY RECORD (DHR) OF MANUFACTURING LOT 1150169 OF CUSA EXCEL 36KHZ CEM NOSECONE WAS REVIEWED. ACCORDING TO THE DHR REVIEW, NO ANOMALIES WERE REPORTED DURING THE PACKAGING PROCESS OF THIS LOT THAT COULD BE RELATED TO THE REPORTED CONDITION. NO SIMILAR COMPLAINTS RELATED TO ¿FOREIGN MATERIAL¿ HAVE BEEN REPORTED FOR THIS LOT 1150169. (B)(4). ONE (1) UNOPENED PACKAGE OF AN UNUSED UNIT FROM CATALOG C6636, CUSA EXCEL 36KHZ CEM NOSECONE CORRESPONDING TO FG LOT # 1150169 WAS RECEIVED FOR EVALUATION IN ITS ORIGINAL PACKAGE. THE UNIT WAS RECEIVED COMPLETELY AND ACCEPTABLY SEALED. UPON INSPECTION, THE FOREIGN MATERIAL WAS OBSERVED ON THE NOSECONE CABLE. DURING THE INSPECTION, IT WAS OBSERVED THAT THE FOREIGN MATERIAL WAS A SMALL LOOSE PARTICULATE COLOR WHITE FOUND ON THE EXTERNAL SIDE OF THE TUBING. THE WHITE PARTICLE WAS COMPARED WITH TAPPI¿S DIRT CHART FINDING IT SIMILAR TO THE TAPPI¿S DIRT DOT WITH DIMENSION .40MM2 WHICH IS BIGGER THAN THE ACCEPTABLE DIMENSION ESTABLISHED IN PROC 3114 AND QCIC 2059 (.10MM2). CONCLUSION: THE FOREIGN MATERIAL ON THE PACKAGING WAS CONFIRMED WITH THE EVALUATION OF THE RETURNED UNIT. HOWEVER, THE ORIGIN OF THE PARTICLE COULD NOT BE DETERMINED. THE ORIGIN OF THIS FOREIGN MATERIAL COULD BE COMING FROM THE FORMING PROCESS OF THE TRAY AT THE SUPPLIER FACILITY OR FROM THE PACKAGING PROCESS, BUT NEITHER OF THEM COULD BE CONFIRMED. THEREFORE, THE ROOT CAUSE FOR THE FOREIGN MATERIAL INSIDE THE PACKAGING TRAY COULD NOT BE DETERMINED AT THIS TIME. RISK CONTROL PROCEDURES ARE IN PLACE TO REDUCE THE PROBABILITY OF OCCURRENCE OF THESE EVENTS. ALTHOUGH NO DEFINITE ROOT CAUSE WAS IDENTIFIED, A POTENTIAL ROOT CAUSE COULD BE THAT THE CLEANING PROCESS MAY HAVE NOT BEEN PERFORMED ADEQUATELY. INTEGRA CONSIDERS THIS COMPLAINT CLOSED. FUTURE INCIDENTS OF THIS NATURE WILL BE DOCUMENTED FOR RECURRENCE AND TRENDING PURPO
AT THE INCOMING INSPECTION OF GOODS, THE DISTRIBUTOR FOUND A PACKAGE WITH FOREIGN MATERIAL INSIDE. NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 592422 | CUSA EXCEL 36KHZ CEM NOSECONE | ULTRASONIC SURGICAL PRODUCTS | LFL | INTEGRA LIFESCIENCES(IRELAND) | 0000001150169 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |