FDA Adverse Event Malfunction Summary report: N

SIG FEM ADPT TORQUE WRENCH

MDR report key: 5060915 · Received September 8, 2015

Report

Report Number
1818910-2015-29925
Event Type
Malfunction
Date Received
September 8, 2015
Date of Event
August 19, 2015
Report Date
August 19, 2015
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE SUBMITTED SIG FEM ADPT TORQUE WRENCH CONFIRMS THE BLACK CAP PROTECTOR IS BROKEN AND HAS DISASSEMBLED FROM THE WRENCH. CAPA (B)(4) WAS INITIATED TO FURTHER INVESTIGATE AND DETERMINE ROOT CAUSE AND CORRECTIVE ACTIONS. (B)(4) HAS BEEN INITIATED TO CHANGE THE DESIGN OF PRODUCT CODE 961673 AS PART OF CAPA (B)(4). THE CURRENT COMPLAINT SAMPLE WAS MANUFACTURED PRIOR TO THE IMPLEMENTATION OF (B)(4). NO FURTHER CORRECTIVE ACTION REQUIRED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4).

Additional Manufacturer Narrative · 1

CONCLUSION AND JUSTIFICATION STATUS FOR MDR: THE DEVICE ASSOCIATED WITH THE REPORTED EVENT WAS NOT RETURNED FOR EVALUATION. EXAMINATION OF A PROVIDED PHOTOGRAPH CONFIRMED THE BLACK PROTECTOR CAP IS MISSING. (B)(4) WAS PREVIOUSLY INITIATED TO FURTHER INVESTIGATE SIG FEM ADPT TORQUE WRENCH PROTECTOR CAP BREAKAGE AND IDENTIFY CORRECTIVE ACTIONS. (B)(4) HAS BEEN INITIATED TO CHANGE THE DESIGN OF PRODUCT CODE 961673 AS PART OF CAPA (B)(4). THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS ABOUT THE ROOT CAUSE OF THE MISSING PROTECTOR CAP WITHOUT THE INSTRUMENT AND PROTECTOR CAP TO EXAMINE. THE CURRENT COMPLAINT SAMPLE WAS MANUFACTURED PRIOR TO THE IMPLEMENTATION OF (B)(4). BASED ON THE INABILITY TO CONCLUSIVELY DETERMINE A ROOT CAUSE, A NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

THE TIP OF THE TORQUE DEVICE CRACKED AND THEN DISSOCIATED FROM HANDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
593783 SIG FEM ADPT TORQUE WRENCH KNEE INSTRUMENT/TRIAL LXH DEPUY ORTHOPAEDICS, INC. SO2012912

Patients

Seq Age Sex Outcome Treatment
1 82 YR