FDA Adverse Event Malfunction Summary report: N

WALKMED INFUSION TRITON FP POLE MOUNT INFUSION PUMP

MDR report key: 5058336 · Received September 4, 2015

Report

Report Number
1723533-2015-00014
Event Type
Malfunction
Date Received
September 4, 2015
Date of Event
August 4, 2015
Report Date
August 6, 2015
Manufacturer
WALKMED INFUSION, LLC.
Product Code
FRN
PMA / PMN Number
K070529
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IN QUESTION WAS RETURNED TO WALKMED INFUSION FOR PRODUCT EVALUATION. THE DEVICE WAS SUBJECTED THROUGH A COMPLETE PRODUCT EVALUATION TESTING AT WHICH POINT IT WAS DISCOVERED THAT THERE WAS AN OPEN STATE OF FREE FLOW WHEN THE IV TUBING IS INSERTED AND DOOR CLOSED. WALKMED HAS PERFORMED A FAILURE INVESTIGATION AND IDENTIFIED NORMAL WEARING COMPONENTS AS THE CAUSE.

Description of Event or Problem · 1

DURING PUMP SET-UP, THE IV WAS NOTED TO BE CONSISTENTLY DRIPPING WITHOUT THE PUMP BEING STARTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
588475 WALKMED INFUSION TRITON FP POLE MOUNT INFUSION PUMP INFUSION PUMP FRN WALKMED INFUSION, LLC.

Patients

Seq Age Sex Outcome Treatment
1 Other