FDA Adverse Event Injury Summary report: N

TEUMO CARDIOVASCULAR PROCEDURE KIT- CONVENIENCE TUBING PACK

MDR report key: 5058012 · Received September 4, 2015

Report

Report Number
1212122-2015-00010
Event Type
Injury
Date Received
September 4, 2015
Date of Event
August 14, 2015
Report Date
September 4, 2015
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS, CORP.
Product Code
OEZ
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT WAS CONFIRMED. A 42 INCH LENGTH OF 1/4 X 1/16 -X-COATED TUBING WAS RETURNED AND A SPLIT OF APPROXIMATELY 0.55 INCH IN LENGTH WAS NOTED ON THE TOP OF THE TUBING. THIS X-COATED TUBING WAS MANUFACTURED FROM A LOT OF TUBING WHERE NO OTHER COMPLAINTS OF SPLITS HAVE OCCURED. SEVERAL TUBING DIMENSIONS WERE ASSESSED AROUND THE SPLIT OF THE TUBING AND ALL DIMENSIONS WERE WITH-IN SPECIFICATIONS. THE DEVICE HISTORY RECORD WAS REVIEWED AND NO ISSUES WERE NOTED. ASSOCIATE AWARENESS TRAINING WILL BE CONDUCTED AS PART OF THE ON-GOING QUALITY PROGRAM. NO FURTHER ACTIONS WILL BE TAKEN AT THIS TIME. (B)(4). (B)(6).

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT DURING CARDIOPULMONARY BYPASS "THE PUMP BOOT RUPTURED". THIS DELAYED THE CASE BY LESS THAN 5 MINUTES AS THEY PROCEEDED TO CHANGE OUT THE PUMP BOOT WITH AND EXTRA ONE THAT WAS AVAILABLE. THE SURGERY WAS COMPLETED SUCCESSFULLY. PER CLINICAL REVIEW: THE PERFUSIONIST WAS ON CPB WITH A 7 KG PATIENT WHEN THE PUMP BOOT RACEWAY RUPTURED. THIS OCCURRED DURING CIRCULATORY ARREST, THEREFORE THEY WEREN'T ACTIVELY PUMPING. IT WAS AT THIS TIME THEY NOTED BLOOD DRIPPING IN THE RACEWAY. THE PERFUSIONIST PROCEEDED TO CHANGE OUT THE PUMP BOOT. THERE WAS NO DELAY TO THE PROCEDURE, NO PATIENT ISSUES, AND THE CASE FINISHED WITHOUT INCIDENCE. THEY DID EXPERIENCE AN APPROXIMATE 50 CC BLOOD LOSS, BUT THIS DIDN'T RESULT IN THE NEED TO TRANSFUSE THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
588644 TEUMO CARDIOVASCULAR PROCEDURE KIT- CONVENIENCE TUBING PACK CARDIOVASCULAR PROCEDURE KIT OEZ TERUMO CARDIOVASCULAR SYSTEMS, CORP. 75170 TF18

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening| R