BLOOD GLUCOSE SYSTEM
Report
- Report Number
- 1052693-2015-01624
- Event Type
- Malfunction
- Date Received
- September 4, 2015
- Date of Event
- August 13, 2015
- Report Date
- December 16, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- OTHER
Narratives
(B)(4) INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: THE USER HAD AN INACCURATE REFERENCE.
(MANUFACTURER NARRATIVE = T, CORRECTED DATA = F) (B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION.
CUSTOMER COMPLAINS OF HIGH RESULTS. CUSTOMER STATES SHE FEELS WELL AND NO MEDICAL ATTENTION IS NEEDED. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 89-109MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 11/14/2017. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED (B)(6) 2015. CUSTOMER PERFORMED A BACK TO BACK BLOOD TEST 180MG/DL AND 168MG/DL. REVIEWED METER MEMORY: 1:194MG/DL (B)(6) 2014 03:49:00 PM FASTING:YES. 2:177MG/DL (B)(6) 2014 01:18:00 AM FASTING:YES . 3:162MG/DL (B)(6) 2014 09:05:00 AM FASTING:YES. 4:129MG/DL (B)(6) 2014 07:06:00 AM FASTING:YES. 5:193MG/DL (B)(6) 2014 04:45:00 AM FASTING:YES. CUSTOMER IS CONCERN WITH THE RESULTS SHE HAVE BEEN GETTING FOR THE PAST FEW DAYS BUT ESPECIALLY TEST TAKEN TODAY (B)(6) 2014; 162/177/194MG/DL. NO ADVERSE EVENT REPORTED.
CUSTOMER COMPLAINS OF HIGH RESULTS. CUSTOMER STATES SHE FEELS WELL AND NO MEDICAL ATTENTION IS NEEDED. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 89-109MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 11/14/2017. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED (B)(6) 2015. CUSTOMER PERFORMED A BACK TO BACK BLOOD TEST 180MG/DL AND 168MG/DL. REVIEWED METER MEMORY: (B)(6). CUSTOMER IS CONCERN WITH THE RESULTS SHE HAVE BEEN GETTING FOR THE PAST FEW DAYS BUT ESPECIALLY TEST TAKEN TODAY (B)(6) 2014; 162/177/194MG/DL. NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 588037 | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PR2167 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |