PEAK PULSAR II GENERATOR

MDR report key: 5056009 · Received September 4, 2015

Report

Report Number
1226420-2015-00147
Event Type
Malfunction
Date Received
September 4, 2015
Date of Event
August 7, 2015
Report Date
October 5, 2015
Manufacturer
MEDTRONIC ADVANCED ENERGY, LLC
Product Code
GEI
PMA / PMN Number
K102029
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT (B)(4). EVALUATION PROCESS: *PERFORMED VISUAL INSPECTION ON UNIT PER (B)(4). - UNIT HAS BARE METAL ON BOTH BOTTOM SIDES OF ENCLOSURE. *PERFORMED BASELINE FUNCTIONAL TESTING PER (B)(4). - NO PERFORMANCE ISSUES FOUND. UNIT DELIVERED CORRECT RF ENERGY IN BOTH CUT <(><<)>(><(>&<)><(><<)>)> COAG. ROOT CAUSE: THE RESISTIVE LOADS USED INSIDE COMMERCIAL OFF-THE-SHELF ANALYZERS ARE LOCATED IN EARTHED METAL ENCLOSURES. THE ENERGY BEING DELIVERED BY THE PULSAR II IS SHUNTED OFF TO THE EARTHED ENCLOSURE IN THE FORM OF RF LEAKAGE CURRENT¿WHICH CAN CONFUSE THE POWER READING OF THE ANALYZER BECAUSE IT TRIES TO MONITOR CURRENT AND VOLTAGE. OUR MEASUREMENT SYSTEM PLACES THE RATED LOAD ON A WOODEN TABLE WHERE IT IS ISOLATED FROM EARTHED METAL. WE USE FLUE TRUE RMS METERS THAT MEASURE ACTUAL HEAT PRODUCED BY THE WAVEFORM. OUR SET-UP IS NOT CONFUSED BY THE COUPLED RF LEAKAGE CURRENT AND GENERATES ACCURATE READINGS. (AS PER (B)(4) ¿ SR. ELECTRICAL ENGINEER). BARE METAL ON ENCLOSURE IS DUE TO ROUGH HANDLING OUT IN THE FIELD. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

BIOMED PERFORMED PREVENTATIVE MAINTENANCE TESTING USING A DNI 454 ESU ANALYZER AND A TNA DEVICE AND FOUND THAT THE OUTPUT ON ALL SETTINGS BELOW COAG 5 WERE HIGHER THAN THE ALLOTTED TOLERANCE.THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

BIOMED PERFORMED PREVENTATIVE MAINTENANCE TESTING USING A DNI 454 ESU ANALYZER AND A TNA DEVICE AND FOUND THAT THE OUTPUT ON ALL SETTINGS BELOW COAG 5 WERE HIGHER THAN THE ALLOTTED TOLERANCE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
588187 PEAK PULSAR II GENERATOR ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI MEDTRONIC ADVANCED ENERGY, LLC PS100-102

Patients

Seq Age Sex Outcome Treatment
1