VERSACARE FRAME
Report
- Report Number
- 1824206-2015-00866
- Event Type
- Malfunction
- Date Received
- September 4, 2015
- Date of Event
- August 14, 2015
- Report Date
- August 14, 2015
- Manufacturer
- HILL-ROM BATESVILLE
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE HILL-ROM TECHNICIAN FOUND THE SIDE COMMUNICATION CABLE WAS THE ISSUE. PER THE HILL-ROM USER MANUAL, WARNING: THE BED EXIT ALARM SYSTEM IS NOT INTENDED AS A SUBSTITUTE FOR GOOD NURSING PRACTICES. THE BED EXIT ALARM SYSTEM MUST BE USED IN CONJUNCTION WITH A SOUND RISK ASSESSMENT AND PROTOCOL. PER THE HILL-ROM USER MANUAL, IF THE BED EXIT ALARM DOES NOT ARM AND ALL THREE MODE INDICATORS ARE FLASHING, REMOVE THE PATIENT AND ZERO THE BED EXIT SYSTEM. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE SIDE COMMUNICATION CABLE TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE BED EXIT ALARM WOULD ALARM, BUT WOULD NOT SEND A CALL TO THE NURSES' STATION. THE BED WAS LOCATED IN ROOM 8404 AT THE FACILITY. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 586801 | VERSACARE FRAME | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM BATESVILLE | P3200K000425 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |