FDA Adverse Event Death Summary report: N

CAPSURE SP NOVUS

MDR report key: 5053802 · Received September 3, 2015

Report

Report Number
2649622-2015-10629
Event Type
Death
Date Received
September 3, 2015
Date of Event
December 22, 2013
Report Date
August 24, 2015
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P830061
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED. IT WAS FOUND THAT THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. IT WAS ALSO NOTED THAT THE PROXIMAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. (B)(4).

Description of Event or Problem · 1

THE LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. ADDITIONAL INFORMATION OBTAINED INDICATED THE LEAD WAS EXPLANTED POST-MORTEM FOR CREMATION. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
583956 CAPSURE SP NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 4092-58

Patients

Seq Age Sex Outcome Treatment
1 00090 YR Death (B)(4)