FDA Adverse Event Summary report: N

76" FLOORLOADER PREVAC STERILIZER

MDR report key: 5053043 · Received September 3, 2015

Report

Report Number
3005899764-2015-00058
Date Received
September 3, 2015
Date of Event
August 11, 2015
Report Date
September 3, 2015
Manufacturer
STERIS MEXICO, S. DE R.L. DE C.V.
Product Code
FLE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN INSPECTED THE STERILIZER AND FOUND THAT THE HEAT EXCHANGER REQUIRED REPLACEMENT. THE TECHNICIAN STATED THAT THE HEAT EXCHANGER WAS LEAKING WATER INTO THE CHAMBER DRAIN CAUSING EXCESS STEAM TO ACCUMULATE. THE TECHNICIAN REPAIRED THE STERILIZER, RAN A TEST CYCLE AND CONFIRMED IT TO BE OPERATING PROPERLY.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT UPON OPENING THE STERILIZER DOOR AFTER A COMPLETED CYCLE EXCESS STEAM EMITTED. THE FIRE ALARM WAS TRIGGERED HOWEVER, THE FIRE DEPARTMENT WAS NOT DISPATCHED AND NO EVACUATIONS OCCURRED. NO REPORT OF INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
585299 76" FLOORLOADER PREVAC STERILIZER STERILIZER FLE STERIS MEXICO, S. DE R.L. DE C.V.

Patients

Seq Age Sex Outcome Treatment
1