FDA Adverse Event
Summary report: N
76" FLOORLOADER PREVAC STERILIZER
MDR report key: 5053043
·
Received September 3, 2015
Report
- Report Number
- 3005899764-2015-00058
- Date Received
- September 3, 2015
- Date of Event
- August 11, 2015
- Report Date
- September 3, 2015
- Manufacturer
- STERIS MEXICO, S. DE R.L. DE C.V.
- Product Code
- FLE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A STERIS SERVICE TECHNICIAN INSPECTED THE STERILIZER AND FOUND THAT THE HEAT EXCHANGER REQUIRED REPLACEMENT. THE TECHNICIAN STATED THAT THE HEAT EXCHANGER WAS LEAKING WATER INTO THE CHAMBER DRAIN CAUSING EXCESS STEAM TO ACCUMULATE. THE TECHNICIAN REPAIRED THE STERILIZER, RAN A TEST CYCLE AND CONFIRMED IT TO BE OPERATING PROPERLY.
Description of Event or Problem · 1
THE USER FACILITY REPORTED THAT UPON OPENING THE STERILIZER DOOR AFTER A COMPLETED CYCLE EXCESS STEAM EMITTED. THE FIRE ALARM WAS TRIGGERED HOWEVER, THE FIRE DEPARTMENT WAS NOT DISPATCHED AND NO EVACUATIONS OCCURRED. NO REPORT OF INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 585299 | 76" FLOORLOADER PREVAC STERILIZER | STERILIZER | FLE | STERIS MEXICO, S. DE R.L. DE C.V. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |