FDA Adverse Event Malfunction Summary report: N

SPACELABS XPREZZON BEDSIDE MONITOR

MDR report key: 5051078 · Received September 2, 2015

Report

Report Number
3010157426-2015-00169
Event Type
Malfunction
Date Received
September 2, 2015
Date of Event
August 6, 2015
Report Date
November 16, 2015
Manufacturer
SPACELABS HEALTHCARE INC.
Product Code
MHX
PMA / PMN Number
K112962
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

SPACELABS HAS LAUNCHED AN INVESTIGATION IN THIS EVENT AND WILL FILE A SUPPLEMENTAL REPORT WHEN THE INVESTIGATION IS COMPLETE.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS REPAIRED ON SITE BY A SPACELABS FIELD SERVICE ENGINEER. THE REPORTED PROBLEM WAS VERIFIED. PART NUMBER 050-0619-02 SERVICE KIT WITH CPU PCBA WAS USED TO REPLACE THE NON-FUNCTIONING DISPLAY COMPONENTS. THE REPAIRED UNIT PASSED ALL FUNCTIONAL TESTS AND WAS RETURNED TO THE CUSTOMER. LACK OF A DISPLAY WAS THE WARNING SIGN THAT THE CUSTOMER NOTICED AND THAT PROMPTED THE ACTION OF REMOVING THE PRODUCT FROM USE. THIS REPORT IS COMPLETE AND THIS PARTICULAR ISSUE IS CONSIDERED CLOSED. SPACELABS MONITORS ITS COMPLAINTS FOR SIMILAR PRODUCT ISSUES TO DETERMINE NEED FOR ADDITIONAL INVESTIGATION.

Description of Event or Problem · 1

SPACELABS RECEIVED A REQUEST FOR SERVICE REPAIR FOR XPREZZON BEDSIDE MONITOR MODEL 91393 THAT LOST DISPLAY DURING USE ON (B)(6) 2015. NO ONE WAS INJURED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
583277 SPACELABS XPREZZON BEDSIDE MONITOR S-CLASS MULTIPARAMETER PATIENT MONITOR MHX SPACELABS HEALTHCARE INC. 91393

Patients

Seq Age Sex Outcome Treatment
1