FDA Adverse Event Summary report: N

RAPIDPOINT 500

MDR report key: 5051060 · Received September 2, 2015

Report

Report Number
1217157-2015-00130
Date Received
September 2, 2015
Date of Event
July 31, 2015
Report Date
October 1, 2015
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS
Product Code
KHP
PMA / PMN Number
K113216
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

BARCODE READER S/N (B)(4) WAS RECEIVED FOR ESCALATION INVESTIGATION. THE BARCODE SUBMITTED IN THE COMPLAINT WAS OF VERY POOR QUALITY SHOWING MULTIPLE PRINT VOIDS. SIEMENS SOFTWARE ENGINEERING TESTED RETURNED BARCODE READERS; SEE BELOW RESULTS: CONNECTED THE EXTERNAL BARCODE SCANNER (MODEL#LS2208, SERIAL #(B)(4)) SENT FROM THE CUSTOMER TO A (B)(4) INSTRUMENT (B)(4). SCANNED ALL THE 16 ORIGINAL BARCODE LABELS (CAME WITH THE SCANNER) FROM THE DIFFERENT POSITIONS ON THE PATIENT DEMOGRAPHICS SCREEN. ALL THE BARCODE LABELS WERE READ SUCCESSFULLY. THE RESULT FROM EACH SCAN WAS CONSISTENT WITH THE ASCII STRING PRINTED BELOW THE BARCODE. REPEATED THE STEP 1 TO 2 FOR THE SECOND SCANNER ((MODEL#LS2208, SERIAL #(B)(4), ALSO SENT FROM THE CUSTOMER) ALL THE BARCODE LABELS WERE READ SUCCESSFULLY. THE RESULT FROM EACH SCAN WAS CONSISTENT WITH THE ASCII STRING PRINTED BELOW THE BARCODE. FROM THE TESTING RESULTS, I DIDN'T FIND ANY PROBLEM WITH THE SCANNERS. NO RESPONSE RECEIVED FROM CUSTOMER ON THEIR INTERNAL BARCODE GRADING SPECIFICATION. THE CUSTOMER COMPLAINT REGARDING BARCODE MISREAD ISSUE IS NOT CONFIRMED.

Additional Manufacturer Narrative · 1

SIEMENS REPRESENTATIVE (CSE) RESCANNED SPARE SAME BARCODE LABEL (AE15072149L), NO DISCREPANCY OBSERVED. CSE USED OTHER PRINTED BARCODE LABEL FROM OTHER SITES AND VERIFIED THAT ANALYZER WAS READING CORRECT BARCODE LABEL. CSE REPLACED A VALIDATED BARCODE GUN AND USED IT TO SCAN FOR 4 DIFFERENT AE BARCODE LABELS, NO DISCREPANCY OBSERVED.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT TWO DIGITS IN PATIENT ID BARCODE READ INCORRECT ON THE ANALYZER. CUSTOMER INDICATED THAT PATIENT ID READ AS (B)(6). THERE WAS NO REPORT OF SERIOUS INJURY DUE TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
582049 RAPIDPOINT 500 RP 500 KHP SIEMENS HEALTHCARE DIAGNOSTICS

Patients

Seq Age Sex Outcome Treatment
1