FDA Adverse Event Other Summary report: N

BMR1900 PH, I. S. I. O. VENOUS RESERVOIR BAG

MDR report key: 5048904 · Received August 27, 2015

Report

Report Number
9680841-2015-00318
Event Type
Other
Date Received
August 27, 2015
Date of Event
July 28, 2015
Report Date
February 28, 2017
Manufacturer
SORIN GROUP ITALIA
Product Code
DTN
PMA / PMN Number
K112771
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

SORIN GROUP (B)(4) MANUFACTURES THE BMR1900 THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH REPORTS IS FILED ON BEHALF OF SORIN GROUP (B)(4). SORIN GROUP RECEIVED A REPORT THAT THE OUTLET CONNECTOR OF A BMR1900 BROKE OFF WHILE REMOVING IT FROM THE BRACKET DURING PRIMING. THERE WAS NO PT INVOLVEMENT. THE INVESTIGATION IS ONGOING. A F/U REPORT WILL BE SENT WHEN THE INVESTIGATION IS COMPLETE.

Additional Manufacturer Narrative · 1

(B)(4). SORIN GROUP (B)(4) MANUFACTURES THE BMR 1900. THE INCIDENT OCCURRED IN (B)(6). (B)(4). THE COMPLAINED UNIT WAS RETURNED TO SORIN GROUP (B)(4) FOR FURTHER INVESTIGATION. VISUAL INSPECTION CONFIRMED THAT THE BLOOD OUTLET CONNECTOR WAS COMPLETELY BROKEN OFF. SIMULATED USE TESTING SUGGESTED THAT THE FORCE TO INDUCE BREAKAGE IS SIGNIFICANTLY HIGHER THAN THE NORMAL FORCE APPLIED TO SET-UP AND TEAR-DOWN THE BAG FORM THE BRACKET. IN ORDER TO IMPROVE THE RESISTANCE OF THE CONNECTORS TO MECHANICAL STRESS, A CAPA WAS INITIATED AND THE LENGTH OF THE PVC TUBING WAS INCREASED BY 4MM. THIS CHANGE PERMITS A REDUCTION OF THE PULL FORCE TO BE APPLIED TO THE CONNECTORS DURING THE SET-UP AND TEAR-DOWN OF THE BAG WITHOUT IMPACTING THE POSITIONING OF THE BAG ONTO THE BRACKET. AN ADDITIONAL CAPA WAS INITIATED TO GUARANTEE CORRECT DISTRIBUTION OF THE SOLVENT BY INTRODUCING A SEMI-AUTOMATIC EQUIPMENT TO AVOID EXCESSIVE DISTRIBUTION OF THE BONDING AGENT. THE INVOLVED UNIT WAS MANUFACTURED BEFORE IMPLEMENTATION OF THE ABOVE CORRECTIVE ACTIONS.

Description of Event or Problem · 1

SORIN GROUP RECEIVED A REPORT THAT THE OUTLET CONNECTOR OF A BMR1900 BROKE OF WHILE REMOVING IT FROM THE BRACKET DURING PRIMING. THERE WAS NO PT INVOLVEMENT.