BMR1900 PH, I. S. I. O. VENOUS RESERVOIR BAG
Report
- Report Number
- 9680841-2015-00318
- Event Type
- Other
- Date Received
- August 27, 2015
- Date of Event
- July 28, 2015
- Report Date
- February 28, 2017
- Manufacturer
- SORIN GROUP ITALIA
- Product Code
- DTN
- PMA / PMN Number
- K112771
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- UNKNOWN
Narratives
SORIN GROUP (B)(4) MANUFACTURES THE BMR1900 THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH REPORTS IS FILED ON BEHALF OF SORIN GROUP (B)(4). SORIN GROUP RECEIVED A REPORT THAT THE OUTLET CONNECTOR OF A BMR1900 BROKE OFF WHILE REMOVING IT FROM THE BRACKET DURING PRIMING. THERE WAS NO PT INVOLVEMENT. THE INVESTIGATION IS ONGOING. A F/U REPORT WILL BE SENT WHEN THE INVESTIGATION IS COMPLETE.
(B)(4). SORIN GROUP (B)(4) MANUFACTURES THE BMR 1900. THE INCIDENT OCCURRED IN (B)(6). (B)(4). THE COMPLAINED UNIT WAS RETURNED TO SORIN GROUP (B)(4) FOR FURTHER INVESTIGATION. VISUAL INSPECTION CONFIRMED THAT THE BLOOD OUTLET CONNECTOR WAS COMPLETELY BROKEN OFF. SIMULATED USE TESTING SUGGESTED THAT THE FORCE TO INDUCE BREAKAGE IS SIGNIFICANTLY HIGHER THAN THE NORMAL FORCE APPLIED TO SET-UP AND TEAR-DOWN THE BAG FORM THE BRACKET. IN ORDER TO IMPROVE THE RESISTANCE OF THE CONNECTORS TO MECHANICAL STRESS, A CAPA WAS INITIATED AND THE LENGTH OF THE PVC TUBING WAS INCREASED BY 4MM. THIS CHANGE PERMITS A REDUCTION OF THE PULL FORCE TO BE APPLIED TO THE CONNECTORS DURING THE SET-UP AND TEAR-DOWN OF THE BAG WITHOUT IMPACTING THE POSITIONING OF THE BAG ONTO THE BRACKET. AN ADDITIONAL CAPA WAS INITIATED TO GUARANTEE CORRECT DISTRIBUTION OF THE SOLVENT BY INTRODUCING A SEMI-AUTOMATIC EQUIPMENT TO AVOID EXCESSIVE DISTRIBUTION OF THE BONDING AGENT. THE INVOLVED UNIT WAS MANUFACTURED BEFORE IMPLEMENTATION OF THE ABOVE CORRECTIVE ACTIONS.
SORIN GROUP RECEIVED A REPORT THAT THE OUTLET CONNECTOR OF A BMR1900 BROKE OF WHILE REMOVING IT FROM THE BRACKET DURING PRIMING. THERE WAS NO PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 570731 | BMR1900 PH, I. S. I. O. VENOUS RESERVOIR BAG | RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS | DTN | SORIN GROUP ITALIA | NA | 1504160088 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |