PRESTO KROGER BGMS
Report
- Report Number
- 3004637226-2015-00031
- Event Type
- Injury
- Date Received
- September 1, 2015
- Date of Event
- August 2, 2015
- Report Date
- September 1, 2015
- Manufacturer
- AGAMATRIX, INC.
- Product Code
- NBW
- PMA / PMN Number
- K073573
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE HAS BEEN REQUESTED AND HAS BEEN RETURNED TO THE MANUFACTURER. CONFIRMATION TESTING SHOWS THE METER TO BE OPERATING WITHIN SPECIFICATION. THE TEST STRIP LOT DHR WAS REFERENCED AND THE LOT CLEARED QC FOR RELEASE. THE OWNER'S GUIDE AND PREVIOUS FIELD RETURNS HAVE BEEN REVIEWED. THERE IS NO EVIDENCE FROM PREVIOUS FIELD RETURNS THE EVENT WAS CAUSED BY A METER MALFUNCTION. BASED ON THE INFORMATION PROVIDED, THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED THOUGH INSULIN, OTHER MEDICATION, AND DIET MAY HAVE BEEN A CAUSE. FACTORS SUCH AS HEMATOCRIT, TEMPERATURE, HUMIDITY, ELEVATION, AND TESTING PROCEDURE MAY HAVE CONTRIBUTED TO A POTENTIALLY HIGHER READING. NO CORRECTIVE ACTION WILL BE PERFORMED BECAUSE NO SYSTEM ERROR CAN BE CONFIRMED. THIS EVENT WILL CONTINUE TO BE TRENDED.
THE REPORTER ADMINISTERED AN INCREASED DOSE OF INSULIN BASED ON THE FASTING MEASUREMENT (255 MG/DL) OF THE KROGER METER. THE REPORTER FELT THE METER READING WAS INDICATIVE OF HIS SYMPTOMS. THE PATIENT INCREASED HIS INSULIN DOSE BASED ON THE MEASUREMENT AND ATE SCRAMBLED EGGS. THE PATIENT LATER PASSED OUT WHILE WAITING AT THE BUS STATION. THE EMTS TOOK HIM TO THE HOSPITAL WHERE HE WAS RELEASED THE NEXT DAY. HE WAS TREATED WITH A GLUCOSE IV AND FELT MUCH BETTER UPON DISCHARGE. A BLOOD-GLUCOSE MEASUREMENT UPON ADMISSION TO THE HOSPITAL WAS NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 577913 | PRESTO KROGER BGMS | BLOOD GLUCOSE METER | NBW | AGAMATRIX, INC. | 8000-04166 | MJ06WN85L |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Hospitalization| O |