POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI
Report
- Report Number
- 1525712-2015-04419
- Event Type
- Malfunction
- Date Received
- August 31, 2015
- Report Date
- January 5, 2016
- Manufacturer
- INVACARE TAYLOR STREET
- Product Code
- ITI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
PRODUCT WAS RETURNED FOR EVALUATION. THE RETURN FIELDS IN ORACLE STATE: CUSTOM POWER WHEELCHAIRS. CONTROLLER/JOYSTICK/OPTIONS, PC BOARD FAILURE. NOT HOLDING TIME/DATE, 3.3VOLTS ON BATTERY. COMPLAINT WAS CONFIRMED. THE UNDERLYING CAUSE COULD NOT BE DETERMINED AFTER REVIEWING THE DOCUMENTATION IN THIS INVESTIGATION.
SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL RECORD WILL BE FILED.
PRODUCT WAS RETURNED FOR EVALUATION. THE RETURN FIELDS IN ORACLE STATE: CUSTOM POWER WHEELCHAIRS. CONTROLLER/JOYSTICK/OPTIONS, PC BOARD FAILURE. NOT HOLDING TIME/DATE, 3.3VOLTS ON BATTERY. DEALER STATES THAT THE JOYSTICK IS CUTTING OUT GOES TO A BLANK SCREEN. AND THE PATIENT STATED THAT IF LEFT ALONE FOR A WHILE IT WILL TURN BACK ON.
DEALER STATES THAT THE JOYSTICK IS CUTTING OUT GOES TO A BLANK SCREEN. AND THE PATIENT STATED THAT IF LEFT ALONE FOR A WHILE IT WILL TURN BACK ON.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 576485 | POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI | WHEELCHAIR, POWERED | ITI | INVACARE TAYLOR STREET | 3GTQ-MCG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |