FDA Adverse Event Malfunction Summary report: N

SIG FEM ADPT TORQUE WRENCH

MDR report key: 5041935 · Received August 31, 2015

Report

Report Number
1818910-2015-29287
Event Type
Malfunction
Date Received
August 31, 2015
Date of Event
August 14, 2015
Report Date
August 26, 2015
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THE INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4).

Additional Manufacturer Narrative · 1

EXAMINATION OF THE SUBMITTED SIG FEM ADPT TORQUE WRENCH CONFIRMS THE BLACK CAP PROTECTOR IS BROKEN AND HAS DISASSEMBLED FROM THE WRENCH. CAPA (B)(4) WAS INITIATED TO FURTHER INVESTIGATE AND DETERMINE ROOT CAUSE AND CORRECTIVE ACTIONS. (B)(4) HAS BEEN INITIATED TO CHANGE THE DESIGN OF PRODUCT CODE 961673 AS PART OF CAPA (B)(4). THE CURRENT COMPLAINT SAMPLE WAS MANUFACTURED PRIOR TO THE IMPLEMENTATION OF (B)(4). NO FURTHER CORRECTIVE ACTION REQUIRED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

THE PLASTIC COATING AROUND THE TORQUE WRENCH IS BROKEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
574414 SIG FEM ADPT TORQUE WRENCH KNEE INSTRUMENT/TRIAL LXH DEPUY ORTHOPAEDICS, INC. SO2012912

Patients

Seq Age Sex Outcome Treatment