FDA Adverse Event
Other
Summary report: N
ADVANTA WEIGH BED
MDR report key: 504151
·
Received January 5, 2004
Report
- Report Number
- 504151
- Event Type
- Other
- Date Received
- January 5, 2004
- Date of Event
- July 1, 2003
- Report Date
- September 1, 2003
- Manufacturer
- HILLROM COMPANY, INC.
- Product Code
- FNJ
- Report Source
- User Facility report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
A PATIENT FELL OUT OF A HILL-ROM ADVANTA BED AND THE BED EXIT ALARM DID NOT SOUND. THE INITIAL INSPECTION BY THE BIOMEDICAL TECHNICIAN FOUND THE BED EXIT ALARM WAS NOT SET. THE BED EXIT ALARM AND THE NURSE CALL TESTED FINE WHEN THE BED EXIT MODE WAS SET CORRECTLY. THE NURSE STATED THAT THE BED AND NURSE CALL ALARMED EARLIER THAT DAY. IT WAS DETERMINED THAT THE NURSE DID NOT RESET THE ALARM AFTER IT HAD BEEN SILENCED. TO TURN THE BED EXIT ON, THE OPERATOR MUST SIMULTANEOUSLY PRESS THE "LOCK" AND THE "MODE" KEY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVANTA WEIGH BED | BED | FNJ | HILLROM COMPANY, INC. | P1600A6037 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Other |