FDA Adverse Event Malfunction Summary report: N

ARCHITECT ANTI-HCV

MDR report key: 5041419 · Received August 30, 2015

Report

Report Number
3002809144-2015-00058
Event Type
Malfunction
Date Received
August 30, 2015
Date of Event
June 1, 2015
Report Date
October 1, 2015
Manufacturer
ABBOTT GERMANY
Product Code
MZO
PMA / PMN Number
P050042
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO CUSTOMER RETURNS WERE AVAILABLE FOR TESTING. A RETAINED KIT OF ARCHITECT (B)(6) REAGENT, LOT NUMBER 52283LI00, WAS TESTED TO EVALUATE SENSITIVITY. ALL CONTROL VALUES MET CONTROL SPECIFICATIONS AND WERE IN THE TYPICAL RANGE. TWO SENSITIVITY PANELS (PANEL A AND PANEL B) AND THE (B)(6) CONTROL WERE TESTED. THE MEAN S/CO OF THE SENSITIVITY PANEL VALUES AND (B)(6) CONTROL VALUES MET SPECIFICATIONS AND ALL GENERATED RESULTS WERE IN THE TYPICAL RANGE AND NO (B)(6) RESULTS WERE OBTAINED. IN ADDITION, THE CLINICAL SENSITIVITY WAS EVALUATED BY TESTING TWO COMMERCIALLY AVAILABLE SEROCONVERSION PANELS (ZEPTOMETRIX (B)(6)) ON ONE INSTRUMENT. THE SEROCONVERSION PANEL RESULTS WERE COMPARED TO ARCHITECT (B)(6) TEST RESULTS PROVIDED BY ZEPTOMETRIX. THE REAGENT DETECTED THE SAME BLEEDS AS REACTIVE WITH COMPARABLE S/CO VALUES FOR THE SEROCONVERSION PANELS. QUALITY METRICS WERE REVIEWED. NO NON-STATISTICAL OR ADVERSE TREND WAS IDENTIFIED FOR THE ISSUE UNDER REVIEW. A REVIEW OF LABELING CONCLUDED THAT THE ISSUE IS ADEQUATELY ADDRESSED. NO MALFUNCTION OR PRODUCT DEFICIENCY WAS IDENTIFIED.

Additional Manufacturer Narrative · 1

(B)(4). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER PROVIDED DATA GENERATED WHILE EVALUATING SPECIFICITY OF THE ARCHITECT ANTI-(B)(4) ASSAY. ONE PATIENT SAMPLE WAS FOUND TO BE (B)(6) WITH THE ARCHITECT (B)(4) ASSAY ON (B)(6) 2015 BUT (B)(6) AT ANOTHER LABORATORY. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
574263 ARCHITECT ANTI-HCV ANTI-HCV MZO ABBOTT GERMANY 52283LI00

Patients

Seq Age Sex Outcome Treatment
1