JUPITER SM OPERATING TABLE
Report
- Report Number
- 9681407-2015-00009
- Event Type
- Malfunction
- Date Received
- August 27, 2015
- Date of Event
- August 4, 2015
- Report Date
- August 4, 2015
- Manufacturer
- TRUMPF MEDIZIN SYSTEME GMBH + CO. HG
- Product Code
- FWX
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
ACCORDING TO THE INVESTIGATION, THE CAUSE OF THE TABLE RUNNING ON ITS OWN WAS DAMAGE TO THE KEYPAD IN THE REMOTE. THIS ISSUE IS DEALT WITH IN THE ONGOING FSCA ((B)(4)). THE KEYPAD USED WAS PRIOR TO THE DESIGN CHANGE. TRUMPF MEDICAL HAS ISSUED SAFETY INSTRUCTIONS TO THE USERS IN THE FIELD SAFETY NOTICE AND THOSE SHOULD BE FOLLOWED UNTIL TRUMPF MEDICAL IS ABLE TO DO THE FINAL EXCHANGE. THE OPERATION TABLE AND THE REMOTE CONTROL WERE CHECKED BY SERVICE AND REPAIRED. ANTICIPATED COMPLETION OF THE FSCA IS JULY 2016.
TRUMPF MEDICAL RECEIVED INFORMATION STATING THAT A JUPITER OR TRUSYSTEM 7500 OPERATING TABLE MOVED INDEPENDENTLY AND WITHOUT THE USER DELIBERATELY PRESSING AN OPERATING ELEMENT. NO PATIENTS OR USERS WERE INJURED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 570597 | JUPITER SM OPERATING TABLE | BATTERY POWERED ELECTROMECHANICAL OPERATING TABLE | FWX | TRUMPF MEDIZIN SYSTEME GMBH + CO. HG | 4003171 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |