STERRAD 100S STERILIZER
Report
- Report Number
- 2084725-2015-00267
- Event Type
- Malfunction
- Date Received
- August 26, 2015
- Date of Event
- July 31, 2015
- Report Date
- July 31, 2015
- Manufacturer
- ADVANCED STERILIZATION PRODUCTS
- Product Code
- MLR
- PMA / PMN Number
- K991999
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). LOAD NOT RECALLED.
ADVANCED STERILIZATION PRODUCTS (ASP) RECEIVED ADDITIONAL INFORMATION FROM THE CUSTOMER STATING THERE WAS NO LOAD INVOLVED WITH THE VENT TIMEOUT ISSUE. AS A RESULT, THIS COMPLAINT IS DEEMED NOT REPORTABLE FOR UNKNOWN LOAD STATUS.
A CUSTOMER REPORTED A VENT TIMEOUT ERROR WITH THEIR STERRAD 100S AND IT IS UNKNOWN IF THE LOAD WAS REPROCESSED PRIOR TO BEING RELEASED FOR USE ON A PATIENT(S). THIS EVENT IS BEING REPORTED AS THE LOAD STATUS IS UNKNOWN. ALTHOUGH THERE IS NO REPORT OF PATIENT INJURY OR HARM AND NO PRIOR INCIDENTS HAVE RESULTED IN SERIOUS INJURY, ADVANCED STERILIZATION PRODUCTS (ASP) HAS DETERMINED THAT THIS SITUATION PRESENTS A POTENTIAL RISK OF INFECTION. ASP RECOGNIZES THAT IN MANY CASES IT WOULD BE DIFFICULT TO TRACE INFECTION BACK TO THE STERILIZATION. AS A MATTER OF POLICY ASP HAS THEREFORE DECIDED TO REPORT CASES OF CANCELLED CYCLES IF THE COMPLAINANT DOES NOT CONFIRM THAT THE LOAD WAS REPROCESSED PRIOR TO USE ON A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 565430 | STERRAD 100S STERILIZER | STERRAD EQUIPMENT (MLR) | MLR | ADVANCED STERILIZATION PRODUCTS | NA | 94050 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |