FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH ULTRA
MDR report key: 503331
·
Received December 8, 2003
Report
- Report Number
- 2939301-2003-09116
- Event Type
- Malfunction
- Date Received
- December 8, 2003
- Report Date
- December 2, 2003
- Manufacturer
- LIFESCAN, INC.
- Product Code
- CFR
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PATIENT
Narratives
Description of Event or Problem · 1
A PATIENT REPORTED BLOOD GLUCOSE RESULTS OF 190, 133, AND 121 MG/DL WITH A LIFESCAN METER, PERFORMED WITHIN 10 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF <=20% AND/0R <=20 MG/DL, THEREBY INDICATING A POTENTIAL MALFUNCTION OF THE METER IN TERMS OF PRECISION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH ULTRA | GLUCOSE MONITORING SYS/KIT | CFR | LIFESCAN, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|